TECHNICAL SPECIFICATIONS
The technical specification details of Comfy Stim® are as follows:
MECHANISM TECHNICAL DESCRIPTION
01
Channel
Dual, isolated between channels
02
Pulse Amplitude
Adjustable, 0-100 mA peak into 500 ohm load each channel.
03
Wave Form
Asymmetrical Bi-Phasic Square Pulse
04
Voltage
0 to 50V (Load: 500 ohm)
05
Power source
One 9 Volt Battery
06
Size
10.1cm(L) x 6.1cm(W) x 2.45cm(H)
07
Weight
150 grams with battery.
08
Pulse Rate
Adjustable, from 2 to 150 Hz, 1 Hz/step
09
Pulse Width
Adjustable, from 50 to 300 microseconds, 10 us/step
10
On Time
Adjustable, 2~90 seconds , 1 Sec./ step
11
Off Time
Adjustable, 0~90 seconds , 1 Sec./ step
12
Ramp Time
Adjustable, 1~8 seconds, 1 Sec./ step, The “On” time will increase and
decrease in the setting value.
13
Mode
Five TENS Modes: B(Burst), N(Normal),M (Modulation),SD1( Strength
Duration), SD2 Two EMS Modes: S(Synchronous), A (Alternate)
14
Burst Mode
Burst rate: Adjustable, 0.5 – 5Hz Pulse width adjustable, 50~300µs
Frequency fixed = 100 Hz
15
Normal Mode
The pulse rate and pulse width are adjustable. It generates continuous
stimulation based on the setting value.
16
Modulation Mode
Modulation mode is a combination of pulse rate and pulse width
modulation. The pulse rate and width are automatically varied in a cycle
pattern. The pulse width is decreased by 50% from its original setting in
0.5 second, then the pulse rate is decreased by 50% from its original
setting in 0.5 second. Total cycle time is 1 second. In this mode, pulse
ate(2-150Hz) and pulse width(50-300µs) are fully adjustable.
17
SD1 Mode
The SD1(Strength-Duration) mode consists of automatic modulation
intensity and pulse width in 40% range. The intensity is always increasing
while the pulse width is decreasing and vice-versa. The intensity is
decreased by 40% while the pulse width is increased by 40% in 5 seconds.
In the next 5 seconds, the intensity is increased by 40% while the pulse
width is decreased by 40%. Total cycle time is 10 seconds. Pulse rate (
2~150Hz) and pulse width (50~300µs) are fully adjustable.