ADJUSTING THE CONTROLS FOR EMS
i. Set On Time
The On Time controls the time of stimulation. By pressing the “Set” control, the
contraction time can be adjusted. Both channels’ stimulation is cycled on and off
by the contraction and relaxation settings. The range is adjustable from
2 seconds to 90 seconds.
As the “ON” time including the ramp up and ramp down time, the setting of it
should be no less than two times of the “Ramp” time. (ON TIME ≥Ramp up +
Ramp down)
j. Set Off Time
The Off Time controls the time of relaxation. By pressing the “SET” control, the
relaxation time can be adjusted. Both channels’ stimulation is cycled on and off
by the contraction and relaxation settings. The range is adjustable from 0
second to 90 seconds.
In Alternate mode, the OFF Time should be equal or more than the ON Time.
(OFF TIME≥ON TIME)