MEASUREMENT OPERATIONS - MEASURING AC CURRENT

Set the measuring function range of AC current

by rotating the dial to 

A

     . The meter will be

within the measuring function range of AC current

as default.

Clamp the instrument to the input current signal

to be tested.

Clamp a single current lead to be tested and

keep it at the bottom of the “U” clamp head.

The clamp meter will automatically select the

proper range and display the true virtual value of

AC current in the centre of the screen.

Frequency response range: 50Hz-60Hz.

Warning

: the largest measuring current

must not be greater than 200A AC during the

measurement of AC current.

MEASURING DC CURRENT

Set the measuring function range of DC current

by rotating the dial to 

A

     . The Meter will be

within the measuring function range of DC current

as default.

Press the 

SELECT

 button to switch into the

measuring function range of DC current.

Clear the base number of the measuring range of

DC current.

As the clamp head sensor reacts to terrestrial

magnetism and surrounding magnetic field, there

generally exists a base number even without

measuring.

Therefore, before measuring DC current, the

base number should be cleared by pressing the

ZERO

 button.

Measurements must be conducted after the base number is cleared and the

measured value returns to zero.

Clamp the instrument to the input current signal to be tested.

Clamp the single current lead to be tested and keep it at the bottom of the “U”

clamp head.

The Meter will automatically select the proper range.

If the current direction is in the same direction as that as that of the clamp head,

it will display the positive value of AC current in the centre of the display screen,

otherwise it will display the negative current value.

Warning

: the largest measuring current must not be greater than 200A DC during

measurement of DC current.

C]C]C] 

Содержание 72-3545

Страница 1: ...1 200A True RMS Fork Meter Models 72 3545...

Страница 2: ...ng AC Voltage Specification 7 Measuring DC Voltage Specification 8 Measuring Resistance Specification 8 Conductivity Testing Specification 8 Diode Measurement Specification 8 Measuring Capacitance Spe...

Страница 3: ...ge to the Meter Use extra care when measuring DC voltage of higher than 42V or 30V AC RMS as there may be danger of an electric shock Do not measure voltage or current higher than the allowable input...

Страница 4: ...nce capacitance and other parameters are measured SELECT BUTTON SELECT function selection button which works at AC DC ACV DCV and RES CNT DIO CAP functions When the function dial is switched to AC DC...

Страница 5: ...nter data hold mode and the Meter will not update the measurement data When you exit reset mode hold state will be cancelled also HOLD the data hold button works with all functions To enter data hold...

Страница 6: ...to be measured into the centre position of the probe Impact resistant strength 1m high impact can be borne Max opening dimension of clamp head Diameter of 14 7mm Max dimension of current wire Diamete...

Страница 7: ...e Resolution ratio Accuracy Overload protection 200 0A 0 1A 2 5 5 200A The DC base number can be cleaned up by pressing the ZERO button AC VOLTAGE SPECIFICATION Measurement range Resolution ratio Accu...

Страница 8: ...circuit voltage is about 1 2V DIODE MEASUREMENT SPECIFICATION Measurement range Resolution ratio Accuracy Overload protection 6 000V 0 001V 0 5V 0 8V 1000V DC 750V AC Open circuit voltage is about 3 3...

Страница 9: ...button to switch into the measuring function range of DC current Clear the base number of the measuring range of DC current As the clamp head sensor reacts to terrestrial magnetism and surrounding ma...

Страница 10: ...be greater than 750V AC MEASURING DC VOLTAGE Insert the black probe into the COM input and insert the red probe into the V input ready for measurement Set the measuring function range by rotating the...

Страница 11: ...ort circuit the meter probe and press the REL button to clear the probe short circuit base number before measurement to eliminate the impact of probe resistance CONDUCTIVITY TESTING Insert the black p...

Страница 12: ...lack probe into the COM input and the red probe into the V input ready for measurement Set the measuring function range by rotating the dial so it aims at the capacitance symbol Press SELECT to switch...

Страница 13: ...al screen will display such four grades as and The greater the induced voltage the higher the grade displayed MAINTENANCE Warning Remove the test probes before uncapping the lid in order to avoid elec...

Страница 14: ...must be treated as Waste Electrical Electronic Equipment WEEE Any WEEE marked products must not be mixed with general household waste but kept separate for the treatment recovery and recycling of the...

Отзывы: