Tenma 72-3545 Скачать руководство пользователя страница 4

4

PRODUCT OVERVIEW

1. Clamp head

 - the sensing device for AC/DC measurement.

Fix the wire into the designated position to measure the

current through the wire.

As the head may touch the live wire at the time of the

measurement, keep your hand below the clamp head.

2. Clamp body

 - hold the middle of the clamp body and do not

exceed the protective baffle at the joining section between the

upper part of the clamp body and the clamp head.

3. Function dial

 - used for the start-up and selection of basic

measurement functions. The salient point on the function dial

indicates the present gear position.

4. Function buttons

 - used for the selection of key functions,

including SELECT, REL/ZERO, HOLD/     and TORCH     .

5. Display screen

 - used to display measurement data and

function symbols. The screen also has the backlight function,

which enables you to measure in low-light environments.

6. Measurement at input terminal

 - it is at the insertion

port of the probe that the voltage, resistance, capacitance and other parameters are

measured.

SELECT BUTTON

SELECT

: function selection button, which works at AC/DC, ACV/DCV and RES/CNT/

DIO/CAP functions.

When the function dial is switched to AC/DC, it defaults to AC function. Press the

SELECT

 button to switch between AC and DC functions.

When the function dial is switched to ACV/DCV it defaults to ACV function. Press

the 

SELECT

 button to switch between ACV and DCV functions.

When the function dial is switched to the RES/CNT/DIO/CAP function it defaults to

the RES function. Press the 

SELECT

 button to switch to the CNT function, press

again to access the DIO function and then again to switch to the CAP function.

REL BUTTON

REL

: measurement button of relative value, which works at other functions except from 

CNT, DIO, NCV and DC functions.

Press 

REL

 under normal measurement mode to switch to relative value measuring

mode.

Under relative value measuring mode, it will show Dn-Df on the display screen

and Df on the sub-display screen, of which Df is the last measured value (relative

value) and Dn is the current measured value.

It will display “REL ∆” at the top left of the display screen, which indicates it has

Содержание 72-3545

Страница 1: ...1 200A True RMS Fork Meter Models 72 3545...

Страница 2: ...ng AC Voltage Specification 7 Measuring DC Voltage Specification 8 Measuring Resistance Specification 8 Conductivity Testing Specification 8 Diode Measurement Specification 8 Measuring Capacitance Spe...

Страница 3: ...ge to the Meter Use extra care when measuring DC voltage of higher than 42V or 30V AC RMS as there may be danger of an electric shock Do not measure voltage or current higher than the allowable input...

Страница 4: ...nce capacitance and other parameters are measured SELECT BUTTON SELECT function selection button which works at AC DC ACV DCV and RES CNT DIO CAP functions When the function dial is switched to AC DC...

Страница 5: ...nter data hold mode and the Meter will not update the measurement data When you exit reset mode hold state will be cancelled also HOLD the data hold button works with all functions To enter data hold...

Страница 6: ...to be measured into the centre position of the probe Impact resistant strength 1m high impact can be borne Max opening dimension of clamp head Diameter of 14 7mm Max dimension of current wire Diamete...

Страница 7: ...e Resolution ratio Accuracy Overload protection 200 0A 0 1A 2 5 5 200A The DC base number can be cleaned up by pressing the ZERO button AC VOLTAGE SPECIFICATION Measurement range Resolution ratio Accu...

Страница 8: ...circuit voltage is about 1 2V DIODE MEASUREMENT SPECIFICATION Measurement range Resolution ratio Accuracy Overload protection 6 000V 0 001V 0 5V 0 8V 1000V DC 750V AC Open circuit voltage is about 3 3...

Страница 9: ...button to switch into the measuring function range of DC current Clear the base number of the measuring range of DC current As the clamp head sensor reacts to terrestrial magnetism and surrounding ma...

Страница 10: ...be greater than 750V AC MEASURING DC VOLTAGE Insert the black probe into the COM input and insert the red probe into the V input ready for measurement Set the measuring function range by rotating the...

Страница 11: ...ort circuit the meter probe and press the REL button to clear the probe short circuit base number before measurement to eliminate the impact of probe resistance CONDUCTIVITY TESTING Insert the black p...

Страница 12: ...lack probe into the COM input and the red probe into the V input ready for measurement Set the measuring function range by rotating the dial so it aims at the capacitance symbol Press SELECT to switch...

Страница 13: ...al screen will display such four grades as and The greater the induced voltage the higher the grade displayed MAINTENANCE Warning Remove the test probes before uncapping the lid in order to avoid elec...

Страница 14: ...must be treated as Waste Electrical Electronic Equipment WEEE Any WEEE marked products must not be mixed with general household waste but kept separate for the treatment recovery and recycling of the...

Отзывы: