MEASURING AC VOLTAGE

Insert the black probe into the “

COM

” input and

the red probe into the “

V Ω

” input.

Set the measuring function range of AC voltage

by rotating the dial to 

V

     .

The Meter will be within the measuring function

range of AC voltage as default.

Connect the voltage to be tested and put the red

and black probes at two sides of the voltage to be

tested. The clamp meter will automatically select

the correct range.

The centre of the screen will display the current

true visual value of AC voltage.

When the voltage value exceeds 30V AC, the

screen will display the high-voltage alarm signal

to warn against a shock hazard.

Note: frequency response scope: 45Hz - 400Hz.

Warning

: AC voltage measuring range must not

be greater than 750V AC.

MEASURING DC VOLTAGE

Insert the black probe into the “

COM

” input and

insert the red probe into the “

V Ω

” input, ready for

measurement.

Set the measuring function range by rotating the

dial to 

V

     .

Press the 

SELECT

 button to switch the Meter

into the measuring function range of DC voltage.

Connect the voltage to be tested and put the red

and black probes at the two sides of voltage to

be tested and the Meter will automatically select

the correct range.

The screen will then display the current true

virtual value of DC voltage.

If the electric potential at the red probe is higher

than that at the black probe, the screen will

display the positive voltage value, otherwise the

negative voltage will be shown.

When the voltage exceeds 42V DC the screen

will display the high-voltage symbol to warn against shock hazard.

Warning

: DC voltage measuring range must not be greater than 1000V DC.

CJCJCJ 

Содержание 72-3545

Страница 1: ...1 200A True RMS Fork Meter Models 72 3545...

Страница 2: ...ng AC Voltage Specification 7 Measuring DC Voltage Specification 8 Measuring Resistance Specification 8 Conductivity Testing Specification 8 Diode Measurement Specification 8 Measuring Capacitance Spe...

Страница 3: ...ge to the Meter Use extra care when measuring DC voltage of higher than 42V or 30V AC RMS as there may be danger of an electric shock Do not measure voltage or current higher than the allowable input...

Страница 4: ...nce capacitance and other parameters are measured SELECT BUTTON SELECT function selection button which works at AC DC ACV DCV and RES CNT DIO CAP functions When the function dial is switched to AC DC...

Страница 5: ...nter data hold mode and the Meter will not update the measurement data When you exit reset mode hold state will be cancelled also HOLD the data hold button works with all functions To enter data hold...

Страница 6: ...to be measured into the centre position of the probe Impact resistant strength 1m high impact can be borne Max opening dimension of clamp head Diameter of 14 7mm Max dimension of current wire Diamete...

Страница 7: ...e Resolution ratio Accuracy Overload protection 200 0A 0 1A 2 5 5 200A The DC base number can be cleaned up by pressing the ZERO button AC VOLTAGE SPECIFICATION Measurement range Resolution ratio Accu...

Страница 8: ...circuit voltage is about 1 2V DIODE MEASUREMENT SPECIFICATION Measurement range Resolution ratio Accuracy Overload protection 6 000V 0 001V 0 5V 0 8V 1000V DC 750V AC Open circuit voltage is about 3 3...

Страница 9: ...button to switch into the measuring function range of DC current Clear the base number of the measuring range of DC current As the clamp head sensor reacts to terrestrial magnetism and surrounding ma...

Страница 10: ...be greater than 750V AC MEASURING DC VOLTAGE Insert the black probe into the COM input and insert the red probe into the V input ready for measurement Set the measuring function range by rotating the...

Страница 11: ...ort circuit the meter probe and press the REL button to clear the probe short circuit base number before measurement to eliminate the impact of probe resistance CONDUCTIVITY TESTING Insert the black p...

Страница 12: ...lack probe into the COM input and the red probe into the V input ready for measurement Set the measuring function range by rotating the dial so it aims at the capacitance symbol Press SELECT to switch...

Страница 13: ...al screen will display such four grades as and The greater the induced voltage the higher the grade displayed MAINTENANCE Warning Remove the test probes before uncapping the lid in order to avoid elec...

Страница 14: ...must be treated as Waste Electrical Electronic Equipment WEEE Any WEEE marked products must not be mixed with general household waste but kept separate for the treatment recovery and recycling of the...

Отзывы: