Chapter 8 Run-Time Alerts
a. USER FUNCTION PARAMETER SETTINGS
The TX Console comes with fixed values for select User function parameter settings. These
values cannot be changed.
Mode
Setting
LOW
MEDIUM
HIGH
IRRIGATION
ON / OFF
CUTTING POWER
(delivery cycle at 100% power)
50ms ON, 50ms OFF
75ms ON, 25ms OFF
Continuous
ASPIRATION
FLOW (cc/min)
10
20
30
VACUUM (mmHg)
100
300
500
b. MESSAGES
PRIMING MESSAGES
DESCRIPTION
Priming ... Please Wait
!
Prime button pressed: will clear when priming cycle is complete.
Vac Test Failed! Prime again
!
Cannot build sufficient vacuum during priming cycle: check the tube set for open
connections or leaks.
Verify no visible kinks in microtip tubing. Deflate inflation cuff by disconnecting
air fitting. Depress the cartridge release button and remove cartridge. Verify
tubing is centered within the two grooves on the cartridge. Reinstall cartridge
and inflation cuff. Press foot pedal. Listen for irrigation pinch valve operation
click. Prime again.
Perform manual priming as follows:
1. Cycle the power switch on the rear of the Console OFF and then
back ON.
2. Select irrigation ON, aspiration HIGH, and cutting power OFF.
3. Depress the foot pedal until the Inflation Cuff is tight around the
irrigation bag and the needle cap is filled with irrigation fluid.
4. Select cutting power LOW, MEDIUM or HIGH and depress the Foot
Pedal to test ultrasonic power.
Fluid Path Blocked! Prime Again
!
Vacuum too high during priming cycle: check tubing setup. Check that fluid is
flowing; check tube set is not pinched or blocked.
Verify no visible kinks in microtip tubing. Deflate inflation cuff by disconnecting
air fitting. Depress the cartridge release button and remove cartridge. Verify
tubing is centered within the two grooves on the cartridge. Reinstall cartridge
and inflation cuff. Press foot pedal. Listen for irrigation pinch valve operation
click. Prime again.
Priming Successful
!
Prime cycle is compete.
RUN TIME MESSAGES
Select Irrigation Settings
!
System is ready for input of IRRIGATION parameter setting.
Select Aspiration Settings
!
System is ready for input of ASPIRATION parameter setting.
Select Cutting Power Settings
!
System is ready for input of CUTTING POWER parameter setting.
Check Handpiece
!
TX MicroTip not plugged in or TX MicroTip failure.
Ensure electrical connector is properly installed. Remove and reinstall.
Check for faulty or damaged microtip.
Confirm foot pedal is properly installed.
Check Cassette
!
Cassette not seated all the way down: should click into place when seated
correctly.