45
Appendix 1
EMC Requirements for TX Console
The TX Console meets all applicable requirements of IEC/EN 60601-1-2 for electromagnetic
compatibility. The TX Console needs special precautions regarding EMC and needs to be
installed and put into service according to the EMC information provided in this Appendix.
Portable and mobile RF communications equipment can affect the operation of the TX Console.
The following is a list of cables that are used with the TX Console that comply with
applicable sections of the EMC Standard:
•
Power Cord
•
TX MicroTip
•
Foot Pedal
Use of cables or accessories other than those specified, with the exception of cables
and accessories sold by the manufacturer of the Tenex Health TX System as replace-
ment parts for internal components, may result in increased EMISSIONS or decreased
IMMUNITY of the Tenex Health TX System.
T
able
5. Guidance and Manufacturer’s Declaration, Electromagnetic Emissions, for the
Tenex Health TX System.
The Tenex Health TX System is intended for use in the electromagnetic environment specified below. The customer
or the user of the Tenex Health TX System should assure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment: guidance
RF emissions
CISPR 11
Group 1
The Tenex Health TX System uses RF energy only for
its internal function. Therefore, its RF emissions are very
low and not likely to cause any interference in nearby
electronic environment.
RF emissions
CISPR 11
Class A
The Tenex Health TX System is suitable for use in all
establishments, including domestic establishments and
those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies