Temperzone SAT-‐3 Installation Setup Guide, Issue 2
Page 4
of 7
Zone 2 temperature
sensor calibration
i
OF
-‐4.0 to +4.0°C
N/A see note 6
Zone 3 temperature
sensor calibration
i
OF
-‐4.0 to +4.0°C
N/A see note 6
Zone 4 temperature
sensor calibration
i
OF
-‐4.0 to +4.0°C
N/A see note 6
Zone 5 temperature
sensor calibration
i
OF
-‐4.0 to +4.0°C
N/A see note 6
Zone 6 temperature
sensor calibration
i
OF
-‐4.0 to +4.0°C
N/A see note 6
Enable/disable zone
control
i
ZONE
En
On or off
Default off. Must be off
to function correctly
without zone control
board.
Maximum difference of
zone setpoint from
zone 1 set point
i
ZONE
SP
2.0 to 10.0°C
N/A see note 6
Minimum number of
zones that must be on
i
ZONE
nn
0, 1 or 2
N/A see note 6
Zone 1 Size
i
1 to 5
N/A see note 6
Zone 2 Size
i
0 to 5
N/A see note 6
Zone 3 Size
i
0 to 5
N/A see note 6
Zone 4 Size
i
0 to 5
N/A see note 6
Zone 5 Size
i
0 to 5
N/A see note 6
Zone 6 Size
i
0 to 5
N/A see note 6
Programmable timer
function
i
tF
1 or 2
Default 1
(standard, see note 5)
Set display backlight
duration
i
bd
0 to 59 s
Default 30s
Set display backlight
intensity
I
bL
0 to 10
Default 10
Set modbus device
address
i
Ad
1 to 99
Default 8
Other settings will not
function correctly.
Set RS485 baud rate
i
br
1 to 4
1:4800
2: 9600
3:14400
Default 4: 19200
Zone controlled auto
mode dead band
i
Ab
0 to 10.0°C
N/A see note 6