SM-BioIIAdvancePlus-EN-1812
7 / 40
2.
REPLACEMENT OF ELECTRICAL ELEMENTS
WARNING
Before any service task ensure to perform a proper decontamination of the cabinet when
applicable as shown in user manual Annexes 5.3 and 5.4.
After finishing any service task, it must be ensured that the cabinet is working properly,
according the checklist shown in chapter 4.
WARNING
Before performing the following service tasks, unplug the power plug of the cabinet in order
to avoid any risk of electrical shock.
It must be ensured the right fixing of the frontal cover by means of the two roods that are
supplied for such purpose.
REMARK
When it is required for the specific service task, the frontal cover must be raised or put back
to the closed position only when the glass is in the closed position. The contacts of the
frontal cover must be pushed back manually.
REMARK
Only use original spare parts from the manufacturer.
2.1.
FUSE OF THE POWER SOCKET
The main power socket has inside protection fuses.
In order to replace these fuses it is necessary to open the cover where they are located, take
the fuses out and replace them by new ones that fulfill the following features:
T10A 250V 5x20mm
.