2.6.
REPLACEMENT OF THE UV LIGHT
For the replacement of the UV light, these steps should be followed:
a)
Unplug the cord plug of the cabinet.
b)
Detach the frontal-glass from the cabinet (by pulling from the lateral handles of the
metal support (Fig.1), and move the glass to the limit upper position.
c)
Take out the UV light from the lamp holders by rotating and pulling down the tube.
(Fig.2).
d)
Assemble the new tube in the lamp-holders, by fully pulling up the tube on the slot of the
lamp-holders, and then rotating it.
Fig.1
Fig.2