provide triggering from
the positive
or
negative-going slope of a wave-
form.
obtained
by
releasing DC and AC
buttons.
enable the sweep to
be triggered
either
internally
from
the vertical
amplifier
or externally.
relate to the coupling of the trigger
circuit.
For very low input frequency
DC should be selected.
2.1.5
VERTICAL
DISPLAY
OFF-ON
release of these buttons,switches off
the
channel
concerned.
If
both
channels are switched off, a straight
line
trace results which
cannot be
shifted
by the POSITION controls,
except when in the X-V mode.
selects triggering from either or both
channels. When alternately triggering
from both channels both INT TRIG
buttons
should
be
released; the
displays
should
be partially
super-
imposed.
provide three display modes for the
vertical channels.
In the CHOP mode, the channels are
alternately switched on and off at a
frequency
of about
150 kHZ; this
mode is suitable at the lower sweep
speeds.
In
the
AL T
mode, each
channel is alternately
displayed for
the duration
of
a
sweep;
this
mode is preferable at higher sweep
speeds.
In the SUM mode, the display is the
addition
of
the
individual
signals.
CHl
POSITION is used to shift the
trace, CH2 POSITION is a Fine shift
in the SUM mode.
If INVERT is depressed,the resultant
display is the difference between the
two input signals.
displaces each trace in the vertical
direction
except when both channels
are off.
In
the
X-V
mode
irrespective of
button settings CH 1 provides a verti-
cal shift and CH2 a horizontal shift.
the setting of this button determines
whether the CH2 signal is displayed in
the same polarity
as the input signal
or inverted. The inverted setting is
used
to
display
the
difference
between two signals in the SUM mode.
provides twelve steps of attenuation
of
each
channel's
input
signal.
Calibrated sensitivities are only valid
when VARIABLE
is fully clockwise.
enables
all
deflection
sensitivities
between that selected by the VOLTS/
DIV switch and the next below to be
covered. The control
must be fully
clockwise for a calibrated display; for
Xl0
gain the knob should be pushed
in.
selects the input signal coupling. In
the DC position, the signal from the
INPUT connector is coupled directly
to the attenuator.
In the
AC position
a capacitor
is
inserted in series.
In the GND position the input to the
attenuator
is grounded and socket
disconnected; this position
enables
the 0 V D.C. level of a trace to be
ascertained.
2.1.6
INPUT AND OUTPUT CONNECTORS
1. INPUTS
BNC
2. OUTPUTS
CAL
connectors are linked to the vertical
channel attenuators via the DC-GND-
AC switch described above.
there
is a BNC connector
in the
sweep section
of
the
front-panel.
This enables either external triggering
signals to be applied, or in the EXT
condition of the VARIABLE
speed it
provides the EXT X input. The con-
nector is D.C. coupled to both trigger
and horizontal amplifier circuits. An
external blocking capacitor may be re-
quired to remove the D.C. component.
Input resistance is 100 kU
A
X5 amplification of the horizontal
display
is obtained
by pulling
out
FINE position. If a dual-trace display
is required on EXT X, the vertical
display mode must be set to CHOP,
the AL T and SUM modes will provide
only one trace.
at thE; rear of the instrument
and
connected via an isolating capacitor
to the CRT cathode. A negative-going
signal is thus necessary to intensify
the trace while a positive-going signal
will blank it.
socket
provides
a
waveform
for
checking the calibration of the vertical
channels.
The repetition
rate
is
at
supply
frequency.
provides a fast-edged positive-going
rectangular
pulse
lasting
for
the
duration of the sweep.
The gate out signal or 0.5 V peak to
peak 1 kHz squarewave is used for
calibrating probes as follows.
1. Connect the probe to INPUT 1.
2. Set
VOLTS/DIV
to
.1 (X10),
10 mV (Xl001.
3. Turn VARIABLE
fUlly clockwise.