E10364-140206
PFC-12
Page 2
1.2.1 CONNECTORS
The reading unit has a power button and 4 connectors:
A power connector for connecting a battery charger
A connector for connecting the removable module to the PFC-12
A PC interface communication connector
A connector for connecting the jumper cable to the vibrating wire sensor.
The two-section connecting cable allows the connection either directly to the vibrating
wire sensor using alligator clips, or to a switching station via a Jaegger connector.
1.2.2 KEYBOARD
The keyboard includes :
4 white arrows : used to move the cursor
‘Enter’ :
used to select or enter information
‘Esc’ :
used to navigate the menu
‘Shift’:
Press 1, 2 or even 3 times on
‘’shift’’ and on the appropriate key to select
letters or symbols (blue)
‘Num Lock’:
used to select the digit.
2 HOW TO USE THE UNIT
Put the unit on. Wait about 30 seconds (the station performs a compilation). Make sure
the removable module is connected to PFC-12.
Connect a vibrating wire sensor to the unit.
Refer to the following color code:
# Jaeger Contact
Signal description
1 (Yellow)
Compensation (if available)
2 (White)
Listen coil
3 (Green)
Sustain coil
4 (Black)
Return signal of the 2 coils