Telegan Tempest 50 Скачать руководство пользователя страница 1

CONTENTS 

 

 

 

 

 

 

Section 

INTRODUCTION 

 

 

 

    1 

 
INSTRUMENT FEATURES   

 

    2 

 

Standard Features 

 

 

    2.1 

 

Options & Upgrades 

 

 

    2.2 

 

Accessories & Consumables 

 

    2.3 

 

Default Settings  

 

     

    2.4 

 
TEMPEST GRAND TOUR 

 

 

    3 

 

Left: Physical Connections 

 

    3.1 

 

Centre: Control Panel   

 

    3.2 

 

Right: Printer & Communications 

    3.3 

 
QUICK START 

 

 

 

    4 

 

Before Starting   

 

     

    4.1 

 

Switching On   

 

 

    4.2 

 

Operation Mode 

 

 

    4.3 

 

Probe Position   

 

 

    4.4 

 

Taking an Emissions Reading   

    4.5 

 

Printing (if option available) 

 

    4.6 

 

Switching Off   

 

 

    4.7 

 
HARDWARE SET UP   

 

     

    5 

 

Power Requirements 

 

 

    5.1 

 

Battery Operation 

 

 

    5.1.1 

 

Mains Adapter   

 

 

    5.1.2 

 

12v DC Adapter 

 

 

    5.1.3 

 

Sample Probe Attachment 

 

    5.2 

 

Printer Roll Insertion   

 

    5.3 

 

Filter Unit 

 

 

 

    5.4 

 

Assembly 

 

 

 

    5.4.1 

 

Filter Replacement 

 

 

    5.4.2 

 

Water Removal  

 

     

    5.4.3 

 

Exhaust Removal 

 

 

    5.5 

 

OPERATION   

 

 

 

    6 

 

Principal Keys   

 

 

    6.1 

 

First Menu Page 

 

 

    6.2 

 

Second Menu Page 

 

 

    6.3 

 
SPECIFICATIONS 

 

 

 

    7 

Содержание Tempest 50

Страница 1: ...1 Switching On 4 2 Operation Mode 4 3 Probe Position 4 4 Taking an Emissions Reading 4 5 Printing if option available 4 6 Switching Off 4 7 HARDWARE SET UP 5 Power Requirements 5 1 Battery Operation...

Страница 2: ......

Страница 3: ...ch panel allows you to achieve the results you need quickly and simply without the risk of moisture or dust damaging the contents As you will see from this manual much of the instrument s operation ca...

Страница 4: ...er trap and particulate filter soft carrying case for instrument and probe 2 2 Options and Upgrades high level CO sensor 20 x concentration range nitric oxide NO sensor and calculated total nitrogen o...

Страница 5: ...al Gas See 2 1 Key 2 in 6 2 Efficiency Basis Gross Nett Key 8 in 6 2 Time UK time Any Key 0 in 6 1 Date DD MM YY MM DD YY Key 0 in 6 1 Options Pressure Units mbar Inches WG Key 2 in 6 3 Print out Inte...

Страница 6: ...llowing reference numbers Ref No Name Description Function 1 Sample Port The port through which the sample gas to be analysed must pass The sample line from the probe should be fitted to the spigot of...

Страница 7: ......

Страница 8: ...he battery symbol indicates that the internal battery is being charged by an external power source 9 Power Indicator An illuminated red LED next to the on off symbol indicates that the unit is switche...

Страница 9: ......

Страница 10: ...and insertion of the printer roll and ribbon 13 Paper Slot The printer paper roll should be fed through the paper slot which also acts as a sharp cutting edge to tear the paper 14 Printer Cover Catch...

Страница 11: ......

Страница 12: ...off keys together 4 3 Operation Mode When switched on the display will read T E L E G A N T E M P E S T 5 0 unless specially configured for you The display will then countdown a stabilising or calibr...

Страница 13: ...tion the probe in the flue to take the required sample 4 5 Taking an Emissions Reading Press key 8 to obtain the emissions data i e the carbon monoxide concentration plus the nitric oxide and calculat...

Страница 14: ...nit is charging whether the unit is operating or not 5 1 2 Mains Adapter the TEMPEST 50 unit can be run continuously from a mains socket using the standard mains charger supplied This adapter should b...

Страница 15: ...d separated at the bottom by the filter inner disc d which allows water to pass through the assembly to the receiving filter bowl b A gasket c is fitted to the filter inner disc to form a seal between...

Страница 16: ......

Страница 17: ...he water level reaches the base of the filter inner disc d and gasket c When draining the filter bowl the sampling pump of the TEMPEST 50 unit should be switched off refer to Section 6 3 6 Pump On Off...

Страница 18: ...characteristic beep will be heard before the start up screen is displayed T E L E G A N T E M P E S T 5 0 Note This may be configured slightly differently in some cases A green light 8 in Fig 2 will i...

Страница 19: ...thermocouples are not connected then the temperature data will not be printed If the optional pressure or nitric oxide NO sensors are not fitted then the respective data will be absent from the print...

Страница 20: ...essing the paper feed function key will advance the paper roll by one line of print and the following message will be displayed P A P E R F E E D 3 Data Output Optional Pressing the data output functi...

Страница 21: ...R C O 0 3 5 C O 2 3 5 N O p p m 1 5 N O x p p m 1 8 E F F N 8 5 5 P r s m B a r 2 8 1 5 4 Backlight Optional Pressing the backlight function key will turn the backlight on and off D I S P L A Y L I G...

Страница 22: ...status reads LOW then the oxygen sensor may require replacement see back page The other line of data is the draught measurement when the pressure sensor is fitted as an option 7 Temperature When the t...

Страница 23: ...he following format E F F 7 3 X A I R O 2 2 0 O 2 1 0 8 C O p p m 5 2 When the efficiency key is pressed again the screen will display the following parameters measured and calculated in the revised f...

Страница 24: ...rameters such as efficiency calculated CO2 etc 6 2 First Menu Page The menu keys enable a further range of TEMPEST 50 functions and parameters to be accessed When prompted for a parameter either a num...

Страница 25: ...be adjusted using the up and down cursor keys Keys 5 and 1 to select the appropriate time interval between printouts As soon as the time interval has been selected by pressing the return key the unit...

Страница 26: ...ould be pressed to choose this parameter 4 View S E T V I E W I N G A N G L E Following this prompt the viewing angle should be adjusted using the up Key 1 and down Key 5 cursor keys to obtain the opt...

Страница 27: ...toggle the intake temperature setting between the on and off states When the desired state is displayed it should be selected by pressing the return key If the yes option is selected then the followin...

Страница 28: ...ation oxygen level to the ambient atmospheric value i e 20 9 v v O2 When this option is chosen the pump will automatically be switched on if not already and a clean air sample should then be drawn thr...

Страница 29: ...nd each pair of numbers represents each field of the date format selected previously After the sixth number has been selected the display will switch to the following time selection screen T I M E 1 1...

Страница 30: ...I C E 6 P U M P The specific options can be chosen by pressing the appropriate numeric key 0 6 and each choice will now be described in detail 1 Pressure ZERO Optional Pressing this key will calibrat...

Страница 31: ...a service code of 000000 to return to the previous screen If an incorrect code is entered reply to any prompts by selecting no using the return key to return to the previous screen 6 Pump On Off Press...

Страница 32: ...charge to full 2 hours Charger input voltage 115v or 230v 50 60 Hz AC Display Liquid crystal display backlight optional Format 4 line 20 character alphanumeric Probe standard construction Pistol grip...

Страница 33: ...rmocouple 0 800o C 32o 1472o F 1o C 0 3 1o C 0 3 2o F Ambient Temperature 0 40o C 32o 104o F 1o C 1o C Pressure Draught 150 mb 0 01mb 0 5 Efficiency Stack loss 0 99 9 0 1 CO2 0 99 9 0 1 The CO and NO...

Страница 34: ...print functions If the unit is frozen then asterisks are printed on both sides of the display Toggles on off 6 Scale Displays the life remaining of the oxygen sensor and the current pressure reading w...