background image

8.  Applications

The ConchaTherm Neptune heated humidification system is designed to heat and humidify 
respiratory gases delivered to adult, pediatric, and infant patients. The following information 
is offered regarding the use of the humidifier in specific applications.

The ConchaTherm Neptune has been tested to operate with ventilators designed to support 
neonatal, pediatric, and adult patients using commonly applied ventilation modalities. These 
ventilators may incorporate some or all of the following performance characteristics:

•   Machine or patient cycled delivery of breath
•   Control of pressure, volume, flow or time

Anesthesia

The ConchaTherm Neptune may be used with non-flammable anesthetic gases with either 
non-heated wire or heated-wire breathing circuits at the minimum recommended flows for 
adults and infants (see Specifications). As flow through the circuit nears the lower limits, the 
use of a heated-wire circuit and a Low-Compliance Concha-Column is recommended in 
order to maintain circuit temperature and avoid a tracking alarm. The humidifier is NOT
RECOMMENDED in anesthesia applications requiring flows of less than those stated in the 
"Specifications". For additional information, see "Low-Flow Applications" later in this section. 

Incubators and Warmers

When setting up the humidifier for use with a radiant warmer or incubator, place the proximal 
patient airway temperature probe away from lamps or other sources of heat so as not to 
affect the temperature reading. 

Warning:

Placing a temperature probe in an environment which is at or near the temper-

ature of the gas, such as an incubator or radiant warmer, may mask a dislodged probe
or cause the heater to shut off, thus reducing the humidity and temperature to the
patient. 

Low-Flow Applications

The ConchaTherm Neptune may be used with a heated-wire breathing circuit at the 
following minimum continuous flows:

•   Adult:

Heated-Wire Circuit - 3 LPM minimum

•   Infant/Pediatric:

Heated-Wire Circuit - 1 LPM minimum

•   Continuous Flow Oxygen: 

Comfort Flo System - 1 LPM minimum

For example, with the humidifier in the adult mode, a volume of three liters of gas must pass 
the proximal patient airway temperature probe every minute while the unit is operating.  In 
low-flow applications, the use of a Low-Compliance Concha-Column is recommended in 
order to maintain circuit temperature and avoid a tracking alarm. When using the humidifier 
in low-flow applications, the warm-up period may be longer due to the slower heat transfer 
through the breathing circuit.

For non-heated wire applications in low flow conditions, the following minimum flows must be 
observed to ensure maintenance of the patient airway temperatures.  If lower flows are 
desired, please use a heated-wire breathing circuit:

•   Adult:

Non-Heated-Wire Circuit - 6 LPM minimum

•   Pediatric:

Non-Heated-Wire Circuit - 5 LPM minimum

•   Neonatal 

Non Heated Wire Circuit – 3 LPM minimum

26

Содержание 425-00

Страница 1: ...ConchaTherm Neptune Heated Humidifier REF 425 00 User s Manual 88924 07 03 07...

Страница 2: ...Characteristics c General System Characteristics 6 Controls and Indicators a Front Panel Indicators b Status Indicators c Front Panel Controls 7 Directions for Use a System Setup b Automatic Operation...

Страница 3: ...temperatures Responsive membrane front panel Internationally accepted Graphic Icon Front Panel LED prompts and audio visual alarms permit rapid detection and correction of problems Temperature trackin...

Страница 4: ...urer Name and Address Do Not Dispose in Landfill Return to Manufacturer for Recycling DRIP PROOF a higher than ORDINARY level of protection from drips leaks and spills for Medical Devices Type BF Devi...

Страница 5: ...m minute volumes recommended in this manual Do not cover the circuit with sheets blankets towels clothing or other materials Do not apply excessive tension to the heated wire harness DO NOT MILK TUBIN...

Страница 6: ...erns within the system associated with normal operation will be disrupted if there is insufficient water to pass through the system causing a high temperature alarm Caution Do not operate the humidifi...

Страница 7: ...d vaporizing the sterile water Ventilatory gas passes through the column where it picks up the moisture in molecular form The heated humidified gas is delivered to the patient through a breathing circ...

Страница 8: ...ce equals the Concha Column s compliance plus the amount of fluid transferred to the reservoir during each inspiratory phase Some fluid in the form of either water vapor or water always transfers duri...

Страница 9: ...a Neptune was designed to provide 100 RH as evidenced by moisture con densation or rainout in the circuit at the recommended flow rates over the settable proximal airway tempera ture settings at zero...

Страница 10: ...esistance to Flow data System Pressures The ConchaTherm circuits are designed to exceed the ISO CD 5367 standard of 60 cm H2O 6 0 3 kPa and are tested to perform at 90 cm H2O System Leakage System lea...

Страница 11: ...romagnetic Emissions and Compatibility This device can be placed in close proximity to other Medical Devices approved to the same standards Communication devices or other equipment emitting RF signals...

Страница 12: ...ation This indicator flashes red when there is a temperature probe problem or an over temperature condition at the patient airway 5 Column Temperature Icon illuminates green when the column button is...

Страница 13: ...m will sound If this hap pens turn the heater off and let it cool If the condition persists take the unit out of service and return to Teleflex Medical for servicing 10 Temperature Gradient Indicator...

Страница 14: ...ay Blue LED between the Sun and Raindrop indicates the actual temperature gradient setting The example below shows a setting of one degree towards the raindrop This setting would increase humidity in...

Страница 15: ...t the ventilator circuit to the Concha Column Remove the yellow cap from the top of the column Connect the humidifier line from the ventilator to one of the ports on the column Connect the inspiratory...

Страница 16: ...he puncture site at the bottom of the reservoir Insert the pin all the way Repeat this procedure for the top puncture pin and puncture site Open the tubing clamp s and gently squeeze the reservoir bot...

Страница 17: ...not turn the humidifier on and do not exit the pause mode until there is regulated gas flow through the system To do so may result in heat buildup causing a bolus of hot air to be delivered to the pat...

Страница 18: ...button to confirm these settings or press and hold the accept button to enter the User Adjustable Operation within 5 minutes the unit will alarm To clear this alarm press the accept button 4 Warm Up...

Страница 19: ...ngs are maintained and automatically re activated when the designated pause time ends To activate the PAUSE Mode press and hold the pause button The timer icon will illuminate and the LED display will...

Страница 20: ...used Warning Do not leave the humidifier ON do not turn the humidifier ON and do not exit the pause mode until there is regulated gas flow through the system To do so may result in heat buildup causin...

Страница 21: ...e new mode press the accept button The unit will return to warm up and then normal operation The selected mode will be retained in memory and the appropriate mode icon will remain illuminated during w...

Страница 22: ...idity RH of the gas to be constant throughout the length of the heated wire circuit Figure 9 shows a typical circuit response and LED display when the airway temperature is set at 37 C and the tempera...

Страница 23: ...chosen should provide a minimum of 30 mg H2O L of delivered gas at 30 C The ConchaTherm Neptune is designed to advise the user of low humidity conditions when the humidifier is in the Invasive mode an...

Страница 24: ...ed As a result less water vapor is created To compensate for this situation the ConchaTherm Neptune features a High Output Mode see figure 12 This mode allows the humidifier to respond to high incomin...

Страница 25: ...ithin the circuit Please note that certain environmental conditions will not allow the humidifier to maintain both the proximal patient airway temperature and the temperature gradient at the levels en...

Страница 26: ...e alarm mute button to silence the cautionary audio alarm The alarm will be silenced for approximately two minutes or until the alarm deactivates When muted the alarm remains activated and the alarm m...

Страница 27: ...ator place the proximal patient airway temperature probe away from lamps or other sources of heat so as not to affect the temperature reading Warning Placing a temperature probe in an environment whic...

Страница 28: ...prior to use Warning This unit requires 120 V nominal 60 Hz sine wave output If used in a transport vehicle requiring an inverter do not use square wave or pulse width modulated sine wave output Over...

Страница 29: ...voir correct low no flow condition Low water level in column or reservoir replace water reservoir and or column B CHECK PROBE Patient and Column Temp Indicators Flash Red Temp probe dislodged or disco...

Страница 30: ...ion before removing the unit from service Note These conditions may not necessarily be accompanied by an alarm All displays on the unit are blank Power cord not connected to power Connect cord to powe...

Страница 31: ...until completion of the Warm up cycle Routine Checks 1 Observe the condensation levels in the breathing circuit with every ventilator check Drain as necessary 2 Replace the Concha Column with each ve...

Страница 32: ...ng operation If a malfunction is detected during a routine check the humidifier will shut off power and display the Service icon See Troubleshooting Check the power cord for damage prior to each use I...

Страница 33: ...lation of the lease Repair of user owned heaters Customer Service shall provide instructions for the return of your heater Upon receipt the heater will be evaluated and an estimate of repair costs sha...

Страница 34: ...coverage or operation of the system as a whole under this warranty GENERAL PROVISIONS THIS WARRANTY IS THE EXCLUSIVE REMEDY AND SETS FORTH THE FULL EXTENT OF TELEFLEX MEDICAL S RESPONSIBILITY THIS WA...

Страница 35: ...ranty inspection and testing of the Product to verify any warranty claim e Product that has the serial number removed or made illegible f Freight and insurance costs to the service center g Scratches...

Страница 36: ...quipment For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation applicable to the frequency of the tran...

Страница 37: ...for power supply lines 1 kV for input output lines 1 kV differential mode 2 kV common mode 5 UT 95 dip in UT for 0 5 cycle 40 UT 60 dip in UT for 5 cycles 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip...

Страница 38: ...ansmitters as determined by an electromagnetic site survey a should be less than the compli ance level in each frequency range b Interference may occur in the vicinity of equipment marked with the fol...

Страница 39: ...00 3 2 Voltage fluctuations flicker emissions IEC 61000 3 3 Compliance Group 1 Class B Class A Complies Electromagnetic environment guidance The CONCHATHERM NEPTUNE Humidifier uses RF energy only for...

Отзывы: