Measure Key
36
OFF
Disables open correction.
MEAS OPEN
Starts open correction (AC).
DCR OPEN
Starts open correction (DC).
Step 4. Press the [MEAS OPEN] or [DCR OPEN]
soft-key. A dialog message,
-circuit the
will appear.
Step 5. Connect the test fixture to the BNC terminals.
Do not connect the DUT to the fixture.
Step 6. Press the [OK] soft-key. The T3LCR will then
measure the open admittances (capacitance
and conductance) at the all test frequency
points.
LCR OPEN
sage is
shown on the display.
When the measurement has finished,
measurement, you can press the [ABORT]
soft-key to abort the OPEN correction.
Step 7. After the OPEN measurement has finished,
the meter will enable OPEN correction
automatically. You can, however, still choose
to manually enable or disable OPEN
correction by depressing the ON or OFF soft-
keys.
The short correction feature of the T3LCR Series
compensates for any residual impedances (R,X) that may
exist outside of calibration, such as the impedance of the
cables and the DUT connection points.
See Figure 4-4 Residual Impedances.
www.
.com
1.800.561.8187
Содержание T3LCR1002
Страница 1: ...www com information itm com 1 800 561 8187 ...
Страница 2: ...www com information itm com 1 800 561 8187 ...
Страница 21: ...T3LCR Series User Manual 19 0V 2 5V Accuracy 0 5 0 005V www com information itm com 1 800 561 8187 ...
Страница 85: ...www com information itm com 1 800 561 8187 ...
Страница 90: ...Examples 88 the capacitors are displayed as shown below www com information itm com 1 800 561 8187 ...
Страница 96: ...Remote Control 94 www com information itm com 1 800 561 8187 ...
Страница 98: ... www com information itm com 1 800 561 8187 ...
Страница 102: ...www com information itm com 1 800 561 8187 ...
Страница 107: ... value MIN MAX www com information itm com 1 800 561 8187 ...
Страница 108: ...www com information itm com 1 800 561 8187 ...
Страница 112: ...www com information itm com 1 800 561 8187 ...
Страница 121: ... ON OFF or 1 0 ON OFF or 1 0 ON OFF or 1 0 value www com information itm com 1 800 561 8187 ...
Страница 135: ...T3LCR Series User Manual 133 264 4 231 291 6 308 6 311 1 www com information itm com 1 800 561 8187 ...