T3LCR Series User Manual
35
The T3LCR
OPEN correction capability cancels
errors due to the stray admittance (G, B) in parallel with
the device under test (Refer to Figure 4-3).
The OPEN correction data is taken at all preset
frequency points, independent of the test
frequency you
set. Refer to the tables on page 130 for the trimming
frequency point lists for each model.
Step 1. Press the [Measure] key.
Step 2. Press the [OPEN SHORT] soft-key.
Step 3. Use the cursor key to select the [OPEN] field.
Soft-key
Function
ON
Enables open correction.
www.
.com
1.800.561.8187
Содержание T3LCR1002
Страница 1: ...www com information itm com 1 800 561 8187 ...
Страница 2: ...www com information itm com 1 800 561 8187 ...
Страница 21: ...T3LCR Series User Manual 19 0V 2 5V Accuracy 0 5 0 005V www com information itm com 1 800 561 8187 ...
Страница 85: ...www com information itm com 1 800 561 8187 ...
Страница 90: ...Examples 88 the capacitors are displayed as shown below www com information itm com 1 800 561 8187 ...
Страница 96: ...Remote Control 94 www com information itm com 1 800 561 8187 ...
Страница 98: ... www com information itm com 1 800 561 8187 ...
Страница 102: ...www com information itm com 1 800 561 8187 ...
Страница 107: ... value MIN MAX www com information itm com 1 800 561 8187 ...
Страница 108: ...www com information itm com 1 800 561 8187 ...
Страница 112: ...www com information itm com 1 800 561 8187 ...
Страница 121: ... ON OFF or 1 0 ON OFF or 1 0 ON OFF or 1 0 value www com information itm com 1 800 561 8187 ...
Страница 135: ...T3LCR Series User Manual 133 264 4 231 291 6 308 6 311 1 www com information itm com 1 800 561 8187 ...