1.5 Electrical
Connections
The pump controller may be placed on top of the pump, or wherever
safety and convenience dictate. Power is supplied to the pump con-
troller through the control cable.
Caution
All connections between the pump and controller should be made
BEFORE the pump is connected to mains power.
1. Connect the pressure transducer cable (which originates from
the top of the pump cylinder) to the nine pin sub-D PRESSURE
TRANSDUCER connector on the pump rear panel (Figure 1-6).
Be sure to tighten the thumbscrews.
2. Connect the control cable (which originates from the first or only
pump rear panel) to the PUMP A connector on the rear panel of
the controller (Figure 1-5) and tighten the thumbscrews. This
cable must be plugged into the PUMP A connector.
NOTE
There are four PUMP connectors on the rear of the controller. Only
the PUMP A connector is wired to supply power to the controller;
therefore, one pump must be attached to this connector.
3. Check the serial number tag to make sure the voltage rating of
the pump is correct.
4. Connect the line cord to the MAINS connector on the back of the
pump.