PROTÉGÉ ZM
PORTABLE, ZERO-MAINTENANCE, SINGLE-GAS MONITOR
SPECIFICATIONS
087-0047, Rev. G/December 2019
A-1
A. SPECIFICATIONS
Table A-1: Monitor Specifications
Category
Specifications
Battery Life
2 years, assuming a maximum 4 minutes of alarm time per day
Alarms
Visual, vibrating, & audible (95 dB)
Tests
Self-test on activation and every 20 hours.
Continuous, automatic battery tests.
Data Log
Lasts 25 events
Housing
Thermoplastic Elastomer (TPE)
Hydrogen
Sulfide
(H
2
S)
Range / Resolution
1 to 100 PPM / 1 PPM
Low Alarm Set Point
10 PPM*
High Alarm Set Point
15 PPM*
Calibration Gas
Concentration
25 PPM
Carbon
Monoxide
(CO)
Range / Resolution
1 to 300 PPM / 1 PPM
Low Alarm Set Point
35 PPM*
High Alarm Set Point
200 PPM*
Calibration Gas
Concentration
100 PPM
Oxygen
(O
2
)
Range / Resolution
1 to 30% volume / 0.1%
Low Alarm Set Point
19.5%*
High Alarm Set Point
23.5%*
Calibration Gas
Concentration
16%
Dimensions
3.7” H X 2.2” W X 1.3” D (94 mm X 56 mm X 33 mm)
Weight
2.7 oz. (76 g)
Intrinsically
Safe Approved
Temperature
Range
H
2
S: -40° F to +122° F (-40° C to +50° C)
CO: -22° F to +122° F (-30° C to +50° C)
O
2
: -58° F to +122° F (-50° C to +50° C)
This differs from the operating temperature range
Operating
Temperature
Range
H
2
S, CO, and O
2
: 14° F to +122° F (-10° C to +50° C)
Outside this temperature range, reduced performance or alarm functionality
may occur.
Operating
Humidity
5% to 95% RH, non-condensing
*Factory defaults can be changed.