ProEM System Manual
Issue 3
•
Detector Temperature (Setup
►
Detector Temperature…)
The default temperature setting is read from the camera. When the array
temperature reaches the set temperature, the Detector Temperature dialog will
report that the temperature is LOCKED. Note that some overshoot may occur.
This could cause temperature lock to be briefly lost and then quickly
re-established. If you are reading the actual temperature reported by the
application software, there may be a small difference between the set and
reported temperature when lock is established. This is normal and does not
indicate a system malfunction. Once lock is established, the temperature will be
stable to within ±0.05°C.
shows the default temperatures for the ProEM:512B and 1024B.
Figure 5-3: Typical WinX/32 Detector Temperature Dialog
NOTE:
The Detector Temperature dialog will not display
temperature information while you are acquiring data.
•
Experiment Setup
►
Main Tab (Acquisition
►
Experiment Setup…)
— Exposure Time: 50 ms
— Accumulations & Number of Images: 1
•
Experiment Setup
►
ROI Tab (Acquisition
►
Experiment Setup…)
Use this function to define the region of interest (ROI).
— Imaging Mode: Select this mode if you are running WinSpec.
— Clicking on Full loads the full size of the chip into the edit boxes.
•
Experiment Setup
►
Timing Tab (Acquisition
►
Experiment Setup…)
— Timing Mode: Free Run
— Shutter Control: Disabled Opened.
NOTE:
In FT mode, the shutter must be disabled open for regular
imaging.
— Safe Mode vs. Fast Mode: Fast Mode
See
4
4
11-
012
6
_0
028
P
RO
EM:512B
P
RO
EM:1024B
Содержание ProEM Series
Страница 1: ...www princetoninstruments com ProEM System Manual 4411 0126 Issue 3 September 26 2019 ...
Страница 10: ...10 ProEM System Manual Issue 3 This page is intentionally blank ...
Страница 16: ...16 ProEM System Manual Issue 3 This page is intentionally blank ...
Страница 30: ...30 ProEM System Manual Issue 3 This page is intentionally blank ...
Страница 50: ...50 ProEM System Manual Issue 3 This page is intentionally blank ...
Страница 88: ...88 ProEM System Manual Issue 3 This page is intentionally blank ...
Страница 114: ...114 ProEM System Manual Issue 3 This page is intentionally blank ...
Страница 122: ...122 ProEM System Manual Issue 3 Figure 7 8 LightField Settings 4411 0126_0078 ...
Страница 136: ...136 ProEM System Manual Issue 3 This page is intentionally blank ...
Страница 146: ...146 ProEM System Manual Issue 3 This page is intentionally blank ...
Страница 152: ...152 ProEM System Manual Issue 3 This page is intentionally blank ...
Страница 161: ...This page is intentionally blank ...