ProEM System Manual
Issue 3
8.1
Acquisition Started but Data Display is Empty
If you have started a data acquisition but no data are appearing in the data display, stop
the acquisition. Look at the Data Display title bar and verify that the first number in the
Region of Interest (ROI) is evenly divisible by 4.
(in 30 x 512 x1) is not evenly divisible by 4.
Figure 8-1: Acquisition Display and Invalid ROI
If the number of horizontal pixels (or super pixels if there is binning in the horizontal) is
not divisible by 4, go to Experiment Setup. On the ROI Setup tab, change the Region of
Interest (ROI) and Binning parameters for your experiment so that the resulting number
of horizontal pixels or super pixels is divisible by 4. Then, re-run the acquisition.
4
4
11
-0
126
_
007
9
Содержание ProEM Series
Страница 1: ...www princetoninstruments com ProEM System Manual 4411 0126 Issue 3 September 26 2019 ...
Страница 10: ...10 ProEM System Manual Issue 3 This page is intentionally blank ...
Страница 16: ...16 ProEM System Manual Issue 3 This page is intentionally blank ...
Страница 30: ...30 ProEM System Manual Issue 3 This page is intentionally blank ...
Страница 50: ...50 ProEM System Manual Issue 3 This page is intentionally blank ...
Страница 88: ...88 ProEM System Manual Issue 3 This page is intentionally blank ...
Страница 114: ...114 ProEM System Manual Issue 3 This page is intentionally blank ...
Страница 122: ...122 ProEM System Manual Issue 3 Figure 7 8 LightField Settings 4411 0126_0078 ...
Страница 136: ...136 ProEM System Manual Issue 3 This page is intentionally blank ...
Страница 146: ...146 ProEM System Manual Issue 3 This page is intentionally blank ...
Страница 152: ...152 ProEM System Manual Issue 3 This page is intentionally blank ...
Страница 161: ...This page is intentionally blank ...