4.3.1
Required Tools
The following tools are required to focus a slit:
•
5/64 Allen wrench
•
#1 Phillips head screwdriver
4.3.2
Focus Procedure
Perform the following procedure to focus the slit:
1.
Illuminate the slit with light from the IsoPlane 81 atomic emission wavelength
calibration lamp.
NOTE:
If you do not have an IsoPlane 81 atomic emission
wavelength calibration lamp, use a suitable atomic emission
lamp.
2.
If necessary, within LightField, on the Experiment Setting tab, open the Regions of
Interest expander and select Bin Center 10 rows. See Figure 4-1.
Figure 4-1:
Configure Regions of Interest
3.
On the Common Acquisition Settings expander, configure an Exposure Time of
50 ms. See Figure 4-2.
Содержание Princeton Instruments IsoPlane 81
Страница 10: ...10 IsoPlane 81 Imaging Spectrometer System Manual Issue 1 This page is intentionally blank ...
Страница 20: ...20 IsoPlane 81 Imaging Spectrometer System Manual Issue 1 This page is intentionally blank ...
Страница 32: ...32 IsoPlane 81 Imaging Spectrometer System Manual Issue 1 This page is intentionally blank ...
Страница 39: ...39 Appendix B Outline Drawings Figure B 1 IsoPlane 81 Outline Drawings ...
Страница 40: ...40 IsoPlane 81 Imaging Spectrometer System Manual Issue 1 This page is intentionally blank ...
Страница 56: ...56 IsoPlane 81 Imaging Spectrometer System Manual Issue 1 This page is intentionally blank ...
Страница 61: ...This page is intentionally blank ...