Page 38 of 42
14.3
Spanning Unit.
The following instructions explain how to use the POWER
POD
-400 command signal to supply a given
channel’s input with the proper span voltage and to set the display to read a transducer’s span value.
You must know four things prior to setting the display’s span value:
1.
You must know the maximum signal level for the specific transducer connected to the
channel to be spanned (5 VDC, 10VDC or 20mA).
2.
You must know the maximum value to be displayed when the maximum signal level for the
transducer is connected to the channel to be spanned.
3.
You must insure that the channel’s multiplier is set to 1.0000.
4.
You must know the channel’s ‘Span Value’. That is, the value that would NOW be
displayed (prior to any adjustment) if the maximum signal was fed into the input of the
channel to be adjusted.
1 and 2. If this information is not on the transducer itself, consult the manufacturer of the
transducer. Their literature or other documentation should specify the maximum output signal
from the transducer and the maximum value to be displayed at that signal level.
3.
Setting the MULTIPLIER to 1.0000.
Press
MODE
Press
3
Press
ENTER
Press
1
,
2
,
3
or
4
depending on which channel is to be adjusted.
Press
ENTER
Press
7
Press
ENTER
Press
4
Press
ENTER
Press
1.0000
using the numeric key pad.
Press
ENTER
You have just set a channel’s multiplier to 1.0000.
4.
If you do not know what the selected channel’s maximum display (Span) value is currently set
to, follow these steps.
Notice that the last step in this exercise is to press ESC, not ENTER.
Press
MODE
Press
3
Press
ENTER
Press
1
,
2
,
3
or
4
depending on which channel is to be adjusted.
Press
ENTER
Press
7
Press
ENTER
Press
3
Press
ENTER
Read the
VALUE
from the display.
Press ESC