Operation PEM-9004
Teledyne Analytical Instruments
40
Note: If both bars extend to the optimal fuel-air relation (the gap
indicated
λ
opt) the firing facility in question is set
correctly.
Press ENTER to close the graphics display.
4.5.1.8
F
UEL
The Fuel menu is used to select a fuel from the list of stored fuels
in memory.
Press the button corresponding to Fuel from the Flue Gas menu.
The following screen appears:
In order to select the new fuel it must appear in the framed box. To
change the current fuel:
•
Use the cycle button to scroll through line by line the available
fuels in memory.
•
When the desired fuel appears in the framed box, press
ENTER or press CLEAR to abort the operation and return the
previous fuel. Pressing either ENTER or CLEAR returns the
display back to the Flue Gas menu.
4.5.1.9
C
ONFIG
See Section 4.4.
4.5.1.10
P
ROGRAM
M
ACROS
Up to three customized measuring configurations can be created
and stored as programmed macros. These macros can be started directly