Section 3 Programming
3-9
• Enter R
INSE
P
RESSURE
time. Pressure
time should be at minimum the entered
vacuum time.
WARNING
During the rinse cycle if the sample enters the
sample chamber and reaches the liquid-sensor
electrodes, the program will stop the vacuum time
and switch to purge.
3.3.5 Liquid Detector and Recycle
The liquid sensor electrodes are mounted at the
top of the sample chamber lid. If the sample
does not touch the electrodes, then S
AMPLE
F
AILED
will appear and R
ECYCLED
will activate
if programmed.The liquid detection feature is
always on. The R
INSE
and R
ECYCLE
option is
located in the C
YCLE
M
ENU
after post purge.
The maximum number of rinse and recycle
cycles is four.