Section 2 Installation
2-11
Figure 2-3 The 4-15 liter bottle rack
2.4.3 12-2.5 Liter
The 12 bottle configuration uses a rack to hold
and align the bottles. To install the bottles into
the sampler:
1. Place the uncapped bottles into the rack.
2. Roll the sequential drive out.
3. Locate the arrow in the center of the rack
and align it with the center block located at
the center of the drive table.
4. Roll the sequential table back into the
refrigerated compartment.
5. Activate the program to R
UN
and the table
will turn to locate #Bottle 1. After the table
has stopped make sure the discharge tube is
centered in the bottle. If the discharge tube
is not centered, rotate/adjust the tube to the
center of the bottle.