Section 5 Maintenance
5-3
5.3 Cleaning the Sample
Chamber
To clean the sample chamber:
1. Stop or Pause the program.
2. Remove the chamber from the sampler.
3. Remove the 10 screws that hold the lid and
chamber body together.
Place the chamber parts into warm soapy
water and clean the chamber and
liquid-sensor electrodes using a brush or
sponge.
4. Assemble the chamber using the screws. Be
sure that the seal is properly seated and the
screws are tight. Loose screws or improperly
placed seal can cause a vacuum leak.
5. Install the sample chamber back into the
unit.
5.3.1 Sample Chamber Removal
To remove the sample chamber:
1. Unscrew the white hose fittings on top of
the sample chamber and loosen the
compression nut. Slide the stainless steel
intake tube in.
2. Use the lever to loosen the set collar.
3. Lift the sample chamber out of the collar.
4. Remove the stainless steel intake tube from
the chamber and the two wires from the
liquid-sensor electrodes.
Содержание Optima