Appendix C: Liquid Cooling Setup
WARNING!
Use of equipment not originally provided by Teledyne
Princeton Instruments for use with liquid cooled cameras
will void any and all warranty coverage of the product.
illustrates the typical hardware configuration for a liquid cooled Kuro
system.
Figure C-1: Block Diagram: Typical Liquid Cooled Kuro Setup Diagram
Perform the following procedure to configure the Kuro system for liquid cooling:
1. Unpack the cooler and hose assembly.
2. Confirm the cooler and hoses are pre-filled with blue-colored coolant.
3. Press one hose connector into its mate on the cooler.
4. Repeat with the second hose connector.
5. Press each hose connector on to its mating connector on the camera as follows:
•
Circulator Out > Camera In
•
Camera Out > Circulator In
Listen for the click.
6. Pull each connector to ensure they are locked.
7. Inspect the set-up to insure the hose connectors are secure at the cooler and
camera.
8. Set both the pump speed and fan speed to level 10 on the front display on the
Circulator.
4
4
11-
015
9
_00
2
0
Power Supply
100/240 V
AC
USB 3.0 Data Cable
KURO
Sample
100/240 V
AC
Coolant
Coolant
Coolant Hoses
Camera
Cooler
Содержание Kuro
Страница 1: ...www princetoninstruments com Kuro sCMOS Camera System Manual 4411 0159 Issue 2 March 28 2019...
Страница 6: ...6 Kuro sCMOS Camera System Manual Issue 2 This page is intentionally blank...
Страница 18: ...18 Kuro sCMOS Camera System Manual Issue 2 This page is intentionally blank...
Страница 24: ...24 Kuro sCMOS Camera System Manual Issue 2 This page is intentionally blank...
Страница 34: ...34 Kuro sCMOS Camera System Manual Issue 2 This page is intentionally blank...
Страница 42: ...42 Kuro sCMOS Camera System Manual Issue 2 This page is intentionally blank...
Страница 47: ...This page is intentionally blank...