HVG-2020B Instruction Manual
EAR99 Technology Subject to Restrictions Contained on the Cover Page
Page 37 of 40
R10
Pirani Gas Bend Factor (Read/Unlock Write)
Pirani gas bend factor. Can only be written to when R8 is set to 7 or 8. Determined using
SETKG command.
R15
Pirani Zero Value (Read/Write)
Pirani zero value in counts
R23
Pirani Temperature Set Point (Read/Factory Write)
Pirani temperature set point for PID control
R24
Pirani Proportional Control (Read/Factory Write)
Pirani drive proportional coefficient
R25
Pirani Derivative Control (Read/ Factory Write)
Pirani drive differential coefficient
R26
Pirani Integral Control (Read/Factory Write)
Pirani drive integral coefficient
R27
Pirani Drive (Read/Factory Only)
Pirani drive counts
R29
Pirani Control Bias (Factory Read/Write)
Lowest Pirani control drive in counts