148-0598 Rev. A 

Page 9 

May  1998 

will automatically advance to the next 
mode in about 5 sec.  
 

5.2.7

  The display will read CAL.H.  

Press and release the push button. The 
display will read C_20 if you have 
previously selected d-4C, or C__1 if you 
have previously selected d-6C. 

 
5.2.8 

With the pressure is exactly 20 

Torr (DV-4) or 1 Torr (DV-6) press and 
release the push button.  This stores the 
tube output in memory and advances to 
the next mode. 

 
5.2.9

  The display will read CAL.L.  

Press and release the push button.  The 
display will read C0.01 (d-4C) or C.001 
(d-6C). 

 
5.2.10

  Evacuate the gauge tube to 

0.01

 

Torr

 if d-4C was selected or to 

0.001

 

Torr

 if d-6C was selected. 

 
5.2.11

  Press and release push button. 

The display will indicate 

0.01 Torr

 if 

d-4C was selected or 

 0.001 Torr 

if d-

6C was selected. This stores the tube 
output in memory and returns the 
MODEL 4/6 to its normal measuring 
mode. 

 
5.2.12

  This completes the custom 

calibration.   
 

5.3 REFERENCE 

TUBE 

CALIBRATION

 

 
The low end of the vacuum range can be 
set with a 

Hastings Reference Tube

such as the 

DB-16D

 which is used with 

the DV-4 range, or the 

DB-20

 is which 

is used with the DV-6 range.  The DB-
16D should have a marked calibration 
value between 0.01 Torr and 0.20 Torr.  
The DB-20 should have a marked 

calibration value between 0.001 Torr and 
0.010 Torr. (1 to 10 millitorr) 
 
You do not need access to the push 
button for this calibration procedure. 
 

5.3.1 

Connect the Reference Tube to 

the AVC module.  Turn the MODEL 4/6 
ON.  You are in the normal measuring 
mode. Wait about 1 minute. 
 

5.3.2

  The display should read the 

pressure marked on the Reference Tube. 
 

NOTE:

  Some DB-16D Reference 

Tubes may be marked in

 mmHg

 and 

some DB-20 Reference Tubes may be 
marked in 

microns Hg

. or 

millitorr

.  To 

convert: 
     1 mmHg        =  

1 Torr

 

     1 micron Hg  =  

0.001 Torr 

     

1 millitorr

      =  0.001 Torr 

 

5.3.3. 

If the displayed pressure does not 

match the pressure marked on the 
Reference Tube, adjust the 
potentiometer marked 

Pmin

 on the top 

of the AVC module until they agree.  Do 
not confuse Pmin with the potentiometer 
marked S.P. also located on the top of 
the AVC module. 

5.3.4.

  This completes the Reference 

Tube calibration. 
 
 

6.0 MAINTENANCE 

 
The only authorized maintenance at the 
user level for the MODEL 4/6 is 
replacement of the ¼ amp fuse in the ac 
line connector. 
 
 

7.0 TROUBLE 

SHOOTING AND 

Содержание HASTINGS 4/6

Страница 1: ...Page 1 May 1998 HASTINGS MOD 4 6 DIGITAL VACUUM INSTRUMENT TELEDYNE BROWN ENGINEERING HASTINGS INSTRUMENTS P O Box 1436 Hampton VA 23661 USA Telephone 757 723 6531 800 950 2468 FAX 757 723 3925 May 1998 148 0598 Rev A ...

Страница 2: ...NTIFY 2 3 0 INSTRUMENT SET UP TUBE INSTALLATION 3 3 1 WIRING 3 3 2 CONFIGURATION 3 3 3 INSTALLING VACUUM GAUGE TUBE 4 4 0 OPERATION 5 4 1 AVC MODULE 5 4 2 TAKING A VACUUM MEASUREMENT 5 4 3 RESOLUTION OF DIGITAL METER 5 5 0 CALIBRATION 5 5 1 RESTORE FACTORY CALIBRATION 5 5 2 CUSTOM CALIBRATION 6 5 3 REFERENCE TUBE CALIBRATION 7 6 0 MAINTENANCE 8 7 0 TROUBLE SHOOTING AND REPAIR 8 7 1 TROUBLE SHOOTIN...

Страница 3: ...ch heats a pair of noble metal thermocouples A change in pressure in the gauge tube changes the thermal conductivity of the gas surrounding the thermocouples thus changing the heat loss of the thermo couples due to thermal conductivity This results in a temperature change in the thermocouples thus changing their dc millivolt output The dc millivolt output varies inversely with pressure with 10mv e...

Страница 4: ...r use Hastings Instruments shall not be liable for any indirect or consequential damages Teledyne Brown Engineering Hastings Instruments will at its option repair replace or refund the selling price of the product if Teledyne Brown Engineering Hastings Instruments determines in good faith that it is defective in materials or workmanship during the warranty period provided that the product is retur...

Страница 5: ...d d 6C To check your current configuration 3 2 1 Plug the AVC module into either a DV 4 or DV 6 tube The tube should be at atmospheric pressure 3 2 2 Remove the front bezel and the front panel Locate the black push button in the lower right hand corner 3 2 3 Turn the MODEL 4 6 ON or plug into 115vac Since the tube is at atmospheric pressure the display should read _or_ for over range It may read E...

Страница 6: ...splay will advance to the next program mode 3 2 13 When the program has advanced to the next mode the display will read either d 4 d 6 d 4C or d 6C NOTE The d 4 and d 6 settings are not used with the AVC module If the tube setting is correct do nothing and the display will automatically advance to the next mode in about 5 sec If you wish to change the tube setting press and release the push button...

Страница 7: ...LUTION OF DIGITAL METER 4 3 1 The resolution of the digital meter varies with pressure For the DV 4 0 10 to 20 Torr range the resolution is Range Resolution 0 010 to 0 100 Torr 0 002 Torr 0 10 to 1 00 Torr 0 01 Torr 1 0 to 10 0 Torr 0 1 Torr 10 to 20 Torr 1 Torr 4 4 3 The resolution of the digital meter varies with pressure For the 0 001 to 2 Torr range the resolution is Range Resolution 0 001 to ...

Страница 8: ... wish to change the tube setting press and release the push button within 5 sec until the correct tube setting is displayed In about 5 sec the display will advance to the next program mode DO NOT PRESS THE MODE SWITCH AGAIN UNTIL THE MODEL 4 6 RETURNS TO THE MEASUREMENT MODE AND READS _or_ This takes about 50 seconds 5 2 CUSTOM CALIBRATION CAUTION The MODEL 4 6 will only accept two high calibratio...

Страница 9: ...e DB 16D which is used with the DV 4 range or the DB 20 is which is used with the DV 6 range The DB 16D should have a marked calibration value between 0 01 Torr and 0 20 Torr The DB 20 should have a marked calibration value between 0 001 Torr and 0 010 Torr 1 to 10 millitorr You do not need access to the push button for this calibration procedure 5 3 1 Connect the Reference Tube to the AVC module ...

Страница 10: ...on when tube was not at the correct pressure REMEDY Re configure MODEL 4 6 or activate Factory Reset 7 2 WARRANTY REPAIR Hastings Instruments will repair or replace at their option any vacuum product covered under the warranty expressed in Section 1 5 that is returned transportation prepaid to their plant in Hampton Virginia All products returned must be accompanied by the Service Form or a writte...

Отзывы: