6.3 Operation of the Serial Interface
Communication with the serial interface of the Model 2002 is via an ASCII data string. In the RS-232
mode the command message consist only of a command string and the terminator. The attention
character and address string are not required, but if they are used they MUST be valid. If all compo-
nents of the ASCII data string are valid the command will be accepted and executed. The RS-232
mode is sometimes referred to as point-to-point mode since only one device may be connected to the
controller at any given time.
A message to the Model 2002 in the RS-485 mode consist of an attention character followed by
the address string , the command string, and the terminator. If all components of the ASCII data
string are valid the command will be accepted and executed. The RS-485 mode is also referred to
as multipoint mode since up to 31 devices may be connected to the same controller in a network
scheme.
COMMAND SYNTAX
In the following examples of syntax codes, the special characters are explained:
The characters in square brackets [ ] represents a command string, either upper or lower case
command characters accepted. All characters must follow each other in the string with no spaces or
other characters.
The characters within wavy brackets { } contain choices for the appropriate command.
The characters within the symbols < > are the common abbreviations for the one digit ASCII
control codes which they represent, (e.g. <CR> represents carriage return).
When entering more than one command in the same data string, they must be separated by a
comma.
All command strings must be followed by the terminator character (carriage return <CR>, also
known as ENTER).
When a lower case character is present in an example it represents an option.
Character Description
Valid Inputs:
a
RS-485 Address (hexadecimal 0-9, A-F)
01 - FF
m
Most Significant Digit Of Mantissa
1 - 9
d
Decimal Digit
0 - 9
e
Exponent
0 - 5
u
Unit Of Pressure
T, M, or P
,
Command Separator (comma)
N/A
<CR>
Command Terminator (carriage return)
N/A
Содержание HASTINGS 2002
Страница 1: ...page 1 MODEL 2002 VACUUM GAUGE HASTINGS INSTRUCTION MANUAL ...
Страница 17: ...page 17 Figure 4 2 MODEL 2002 Analog Output VDC volts vs Pressure Torr ...
Страница 18: ...page 18 ...
Страница 26: ...page 26 ...
Страница 38: ...page 38 ...
Страница 39: ...page 39 Diagrams and Drawings SECTION 10 ...
Страница 44: ...page 44 ...
Страница 45: ...page 45 ...
Страница 46: ...page 46 ...
Страница 47: ...page 47 ...
Страница 48: ...page 48 ...
Страница 49: ...page 49 ...