Operator’s Manual
3
Introduction
The CP030, CP030A, CP031, and CP031A current probes are designed for easy, highly
accurate current measurements. The compact probes offer wide bandwidth with
over-current protection. The probes utilize a combination of Hall-effect and transformer
technology which enables measurements to be made on DC, AC and impulse currents.
The CP030 and CP030A have a 50 MHz bandwidth and are designed to measure
continuous currents up to 30 Amp. The CP030-3M probe has a 10 MHz bandwidth and
is designed to measure continuous currents up to 30 Amp.
NOTE:
All CP030A probes, regardless of cable length, will indicate default probe
bandwidth of 50 MHz in the Attributes section of the Probe dialog.
The CP031 and CP031A have a 100 MHz bandwidth and are designed to measure
continuous currents up to 30 Amp.
The CP03x probes are compatible with any Teledyne LeCroy oscilloscope with a ProBus
interface running Windows XP Professional
®
or later.
The CP03xA probes are compatible with most Teledyne LeCroy oscilloscopes with a
ProBus interface running X-Stream™ firmware version 7.8.x.x or later. See table below.
Compatible
Compatible with Upgrade
Not Compatible
LabMaster 9MZi-A
DDA/SDA/WaveMaster 8 Zi/Zi-A/Zi-B
DDA/SDA/WavePro 7 Zi/Zi-A
HRO/WaveRunner 6 Zi
WaveRunner Xi/Xi-A/Mxi/Mxi-A
HDO8000, HDO6000, HDO4000
MDA800
WaveSurfer 10
WaveSurfer MXs/MXs-A/MXs-B
WaveSurfer Xs-A
WavePro 7000/7000A*
WaveSurfer Xs
†
LabMaster 10 Zi/Zi-A
LabMaster 9 SZi-A
SDA13000, SDA18000
SDA9000, SDA11000
WaveExpert
WaveRunner 6000/6000A
WaveSurfer 400
WaveSurfer 3000
All other legacy instruments
*Compatible if running Windows XP Professional and X-Stream 7.8.x.x or later. †Must be upgraded to Windows XP
Professional,
not
Windows XP Embedded. Contact your regional service center regarding upgrades.
With the ProBus interface, the probe becomes an integral part of the measuring
instrument. The bandwidth limit, Auto Zero and Degauss functions are all controlled
from the instrument’s touch screen user interface. The interface provides power to the
probe, so no external power supply is needed.