The instrument operates on any full scale range from 5 to 5,000 ppm. The range is the
concentration value that equals the maximum voltage or current output on the rear panel of the
instrument.
If the range you select is between 5 and 500 ppm the front panel will read the concentration
anywhere from 0 to 500 ppm regardless of the range selected. If the range selected is from 501
to 5,000 ppm the front panel will read from 0 to 5,000 ppm. The apparently wider range of
front panel readouts is because the M9110AH has 2 internal hardware ranges, namely 0-500
ppm and
0-5,000 ppm. The analog output is scaled for the range selected, however the front panel
reading can display correct concentrations over the entire physical range.
NOTE
Only one of the following range choices can be active at any one time.
There are 3 range choices:
1. Single Range
2. Auto Range
A. Remote Range
3. Independent Ranges
5.3.4.1 Single Range
This range option selects a single range for all output channels (NO, NO
x
, NO
2
) of the
M9110AH. To select Single Range press SETUP-RNGE-MODE-SING, then press ENTR. To set
the value for the range press SETUP-RNGE-SET, and enter the full scale range desired from 5
ppm to 5,000 ppm.
5.3.4.2 Auto Range and Remote Range
Auto Range allows each of the 3 output signals (NO, NO
x
, NO
2
) to automatically range between
a low value and a higher value. The Hi range mode is signaled by a bit on the STATUS
connector see Table 5-8. When the instrument output increases to 98% of the low range value it
will Auto Range into Hi range. In Hi range, when the output decreases to 75% of low range it
will change to the lower range. If you select a Hi range that is less than Low range, the
M9110AH will remain locked in Low range and behave as a Single Range instrument.
Содержание 9110AH
Страница 2: ......
Страница 11: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F x INTENTIONALLY BLANK ...
Страница 13: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 1 2 INTENTIONALLY BLANK ...
Страница 17: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 2 4 Figure 2 2 Rear Panel ...
Страница 23: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 2 9 Figure 2 4 Front Panel ...
Страница 24: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 2 11 ...
Страница 25: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 2 12 Figure 2 5 Assembly Layout ...
Страница 28: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 2 15 INTENTIONALLY BLANK ...
Страница 32: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 3 4 INTENTIONALLY BLANK ...
Страница 35: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 4 3 Figure 4 1 Block Diagram ...
Страница 41: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 4 9 Figure 4 2 External Pump Pack ...
Страница 44: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 5 2 Figure 5 1 Sample Menu Tree ...
Страница 45: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 5 3 Figure 5 2 Setup Menu Tree ...
Страница 88: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 6 6 INTENTIONALLY BLANK ...
Страница 102: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 7 14 INTENTIONALLY BLANK ...
Страница 105: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 8 3 Figure 8 1 Replacing the Particulate Filter ...
Страница 107: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 8 5 Figure 8 2 Converter Assembly ...
Страница 110: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 8 8 Figure 8 3 Reaction Cell Assembly ...
Страница 111: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 8 9 Figure 8 4 Sample Bypass Flow Control Assembly ...
Страница 113: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 8 11 Figure 8 5 Pneumatic Diagram ...
Страница 114: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 8 12 Figure 8 6 Pneumatic Diagram with Zero Span Valves ...
Страница 139: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 9 23 Figure 9 1 Span Calibration Voltage ...
Страница 149: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 9 33 Figure 9 2 CPU Board Jumper Settings ...
Страница 151: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 9 35 Figure 9 3 RS 232 Pin Assignments ...
Страница 158: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 9 42 Figure 9 4 V F Board Jumper Settings ...
Страница 163: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 9 47 Figure 9 5 Power Supply Module Layout ...
Страница 164: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 9 48 Figure 9 6 Electrical Block Diagram ...
Страница 168: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 9 52 Figure 9 7 Ozone Generator Subsystem ...
Страница 171: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 9 55 Figure 9 8 Flow Pressure Sensor ...
Страница 172: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 9 56 Figure 9 9 NOx Sensor Module ...
Страница 173: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 9 57 Figure 9 10 NOx Sensor Module ...
Страница 176: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 9 60 Figure 9 11 PMT Cooler Subsystem ...
Страница 178: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 9 62 Figure 9 12 High Voltage Power Supply ...
Страница 181: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 9 65 INTENTIONALLY BLANK ...