TELEDYNE Model 9110AH NO
X
Analyzer Operator Manual, 01620, Rev. F
9-7
Table 9-1: Test Functions (Continued)
Test Function
Factory Set-Up
Comment
CONV TEMP
O3KL TEMP
700 C
10 C
315 C
5 C
200
10 C
The CPU controls the converter temp to 700 C. After cold start it
requires about 30 min to come to temperature. After temperature is
reached, it should not vary more than 10 C. See Section 9.3.4 for
troubleshooting. Warning limits are < 690 C and > 720 C.
The M9110AH can be equipped with a molybdenum converter as an
option. The troubleshooting is the same as listed above except the
warning limits are <290 C and > 320 C.
If the M9110AH is equipped with the M501 external converter option,
the converter housing is equipped with a heater to warm the ozone
killer catalyst to 200 C.
RCEL PRESS
3 – 10 “-Hg-A
(Thomas 607)
RCEL is the pressure in the Reaction Cell. The instrument is very
sensitive to variations in reaction cell pressure. 10% change in output
per 1"Hg pressure is typical. Higher altitudes will decrease cell
pressure due to lower pump back pressure. Pressures higher than the
acceptable range will decrease instrument sensitivity and increase
noise.
SAMP PRESS
29.5"Hg at sea
level
The sample pressure is taken upstream of the reaction cell. It usually
runs about 0.5" less than ambient pressure due to the restrictions in
the sample intake tubing. Sample pressure should be within 1"Hg of
atmospheric pressure. The pressure sensor used reports absolute
pressure and therefore is sensitive to altitude, weather fronts, and
room air conditioning. Change due to altitude is about 1" per
1000 ft., other changes are 0.4" maximum. Pressurizing the sample
inlet will cause the M9110AH to be noisy and to shift its reading.
SLOPE
1.0 0.3
NO and NO
x
SLOPE. The slope is a software gain factor used to
calibrate the analyzer. When the instrument is span calibrated, this
parameter is set so the analyzer front panel and analog output show
the expected span gas concentration. There can be 2 values for
SLOPE if AutoRange is enabled.
OFFSET
0 25
NO and NO
x
OFFSET. The offset is used to zero the analyzer. When
the instrument is zeroed, the offset term is set so the concentration
reads 0 ppm. There can be 2 values for OFFSET if AutoRange is
enabled.
TIME
This is the time of day clock readout. It is used to time the AutoCal
cycles. The speed of the clock can be adjusted by the CLOCK_ADJ
variable in the VARS menu. The clock can be set via SETUP-
CLOCK-TIME from the front panel.
Содержание 9110AH
Страница 2: ......
Страница 11: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F x INTENTIONALLY BLANK ...
Страница 13: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 1 2 INTENTIONALLY BLANK ...
Страница 17: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 2 4 Figure 2 2 Rear Panel ...
Страница 23: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 2 9 Figure 2 4 Front Panel ...
Страница 24: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 2 11 ...
Страница 25: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 2 12 Figure 2 5 Assembly Layout ...
Страница 28: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 2 15 INTENTIONALLY BLANK ...
Страница 32: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 3 4 INTENTIONALLY BLANK ...
Страница 35: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 4 3 Figure 4 1 Block Diagram ...
Страница 41: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 4 9 Figure 4 2 External Pump Pack ...
Страница 44: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 5 2 Figure 5 1 Sample Menu Tree ...
Страница 45: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 5 3 Figure 5 2 Setup Menu Tree ...
Страница 88: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 6 6 INTENTIONALLY BLANK ...
Страница 102: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 7 14 INTENTIONALLY BLANK ...
Страница 105: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 8 3 Figure 8 1 Replacing the Particulate Filter ...
Страница 107: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 8 5 Figure 8 2 Converter Assembly ...
Страница 110: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 8 8 Figure 8 3 Reaction Cell Assembly ...
Страница 111: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 8 9 Figure 8 4 Sample Bypass Flow Control Assembly ...
Страница 113: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 8 11 Figure 8 5 Pneumatic Diagram ...
Страница 114: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 8 12 Figure 8 6 Pneumatic Diagram with Zero Span Valves ...
Страница 139: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 9 23 Figure 9 1 Span Calibration Voltage ...
Страница 149: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 9 33 Figure 9 2 CPU Board Jumper Settings ...
Страница 151: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 9 35 Figure 9 3 RS 232 Pin Assignments ...
Страница 158: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 9 42 Figure 9 4 V F Board Jumper Settings ...
Страница 163: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 9 47 Figure 9 5 Power Supply Module Layout ...
Страница 164: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 9 48 Figure 9 6 Electrical Block Diagram ...
Страница 168: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 9 52 Figure 9 7 Ozone Generator Subsystem ...
Страница 171: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 9 55 Figure 9 8 Flow Pressure Sensor ...
Страница 172: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 9 56 Figure 9 9 NOx Sensor Module ...
Страница 173: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 9 57 Figure 9 10 NOx Sensor Module ...
Страница 176: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 9 60 Figure 9 11 PMT Cooler Subsystem ...
Страница 178: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 9 62 Figure 9 12 High Voltage Power Supply ...
Страница 181: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 9 65 INTENTIONALLY BLANK ...