TELEDYNE Model 9110AH NO
X
Analyzer Operator Manual, 01620, Rev. F
9-36
9.3.2.1
RS-232 Port
If the RS-232 port is not working, check the following:
Physical Wiring
Refer to Figure 9-3 for the wiring diagram of the DB-9 plug on the M200A rear panel. There are
2 features that make connecting the wiring easier. First is the red/green LED’s on the rear panel.
One device (such as the M9110AH) provides the power to run the red LED, the other device
will provide the power for the green LED. If the wiring is hooked up correctly both LED’s will
be illuminated. There is a DTE-DCE switch, this switch interchanges pin 2 & 3 on the DB-9
connector. Set the DTE-DCE switch so that both LED’s are illuminated.
RS-232 Protocol (BAUD rate, Data bits, Parity)
The communication protocol for each instrument must match. In Figure 9-3 the default RS-232
parameters are listed. The BAUD rate can be changed in the software menus under SETUP-
MORE-COMM-BAUD.
Data Communications Software for a PC
You will need to purchase a software package so your computer can transmit and receive on its
serial port. There are many such programs, we use PROCOMM at TELEDYNE. Once you set
up the variables in PROCOMM and your wiring connections are correct, you will be able to
communicate with the analyzer.
If connecting to a modem, check the following:
Modems are especially difficult because they may have pins that need to be at certain EIA RS-
232 levels before the modem will transmit data. The most common requirement is the Ready to
Send (RTS) signal must be at logic high (+5V to +15V) before the modem will transmit. The
TELEDYNE analyzer sets pin 8 (RTS) to 10 volts to enable modem transmission.
To troubleshoot a modem connection first disconnect the RS-232 cable from the Analyzer and
verify (use a DVM) that you are getting a signal on Pin 2 of the RS232 port on the Analyzer.
The signal will be between -5V and -15V with respect to signal ground (pin 5). If not, there is a
problem with the CPU board or the cable. This is the transmit (TD) signal out of the Analyzer.
This should then be connected to TD input on the modem, normally Pin 2. You may need to
change the DTE/DCE switch on the M9110AH so the signal is on Pin 2 of the modem.
Go to the cable connected to the modem/terminal and verify (use a DVM) that you are getting a
-5V to -15V signal on Pin 3 of the cable. This pin should be connected to Pin 3 of the
TELEDYNE Analyzer.
Check that the voltage level on Pin 8 of the Analyzer is b5V and +15V. This pin
should be connected (through the cable) to Pin 4 of the modem.
Содержание 9110AH
Страница 2: ......
Страница 11: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F x INTENTIONALLY BLANK ...
Страница 13: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 1 2 INTENTIONALLY BLANK ...
Страница 17: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 2 4 Figure 2 2 Rear Panel ...
Страница 23: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 2 9 Figure 2 4 Front Panel ...
Страница 24: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 2 11 ...
Страница 25: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 2 12 Figure 2 5 Assembly Layout ...
Страница 28: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 2 15 INTENTIONALLY BLANK ...
Страница 32: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 3 4 INTENTIONALLY BLANK ...
Страница 35: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 4 3 Figure 4 1 Block Diagram ...
Страница 41: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 4 9 Figure 4 2 External Pump Pack ...
Страница 44: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 5 2 Figure 5 1 Sample Menu Tree ...
Страница 45: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 5 3 Figure 5 2 Setup Menu Tree ...
Страница 88: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 6 6 INTENTIONALLY BLANK ...
Страница 102: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 7 14 INTENTIONALLY BLANK ...
Страница 105: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 8 3 Figure 8 1 Replacing the Particulate Filter ...
Страница 107: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 8 5 Figure 8 2 Converter Assembly ...
Страница 110: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 8 8 Figure 8 3 Reaction Cell Assembly ...
Страница 111: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 8 9 Figure 8 4 Sample Bypass Flow Control Assembly ...
Страница 113: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 8 11 Figure 8 5 Pneumatic Diagram ...
Страница 114: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 8 12 Figure 8 6 Pneumatic Diagram with Zero Span Valves ...
Страница 139: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 9 23 Figure 9 1 Span Calibration Voltage ...
Страница 149: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 9 33 Figure 9 2 CPU Board Jumper Settings ...
Страница 151: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 9 35 Figure 9 3 RS 232 Pin Assignments ...
Страница 158: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 9 42 Figure 9 4 V F Board Jumper Settings ...
Страница 163: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 9 47 Figure 9 5 Power Supply Module Layout ...
Страница 164: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 9 48 Figure 9 6 Electrical Block Diagram ...
Страница 168: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 9 52 Figure 9 7 Ozone Generator Subsystem ...
Страница 171: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 9 55 Figure 9 8 Flow Pressure Sensor ...
Страница 172: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 9 56 Figure 9 9 NOx Sensor Module ...
Страница 173: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 9 57 Figure 9 10 NOx Sensor Module ...
Страница 176: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 9 60 Figure 9 11 PMT Cooler Subsystem ...
Страница 178: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 9 62 Figure 9 12 High Voltage Power Supply ...
Страница 181: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 9 65 INTENTIONALLY BLANK ...