Teledyne API – Model 430 Compact Ozone Analyzer
Maintenance and Service
08311 DCN7342
25
Table 4-3. Ozone Bench Status Byte (FW rev 1.06, not logged, displayed as S:xx during errors)
Bit
Definition
Front Panel Message
0
Data stale bit, no new data, 1 = stale
1
Global error bit, 1 = error pending
2
Startup mode bit, 1 = sensor in startup mode
STARTUP
3
Reading Invalid (Lamp Low Err) | (Negative Reading < -10ppb) | ( DATA INVALID
4
A/D locked
BENCH ADC ERROR
5
N/A
6
Pressure < 9 psia or > 14.9 psia
PRESSURE ERROR
7
N/A
Table 4-4. Ozone Bench Error Byte (FW rev 1.06, logged, displayed as E1:xx during errors)
Bit Description
E/W Conditions
Front Panel Message
0
EEProm Initialized
W
EEProm Initialized
BENCH HW ERROR
1
Lamp Temp Warning
W
50 < Lamp Temp < 62 (deg C)
LAMP TMP WARNING
2
Lamp Temp Error
E
-20 < Lamp Temp < 120 (degC)
(Heater Off)
LAMP ERROR
3
Sample Temp Warning
W
0 < Sample Temp < 55 (deg C)
SAMPLE TMP WARN
4
Lamp Low Warning
W
Reference_mv < 375
LAMP LOW WARNING
5
Lamp Low Error
E
Reference_mv < 250
LAMP ERROR
6
Pressure Warning
W
9 < Pressure < 14.9 (psia)
PRESSURE ERROR
7
Lamp Stability
E
IIR threshold exceeded
LAMP ERROR
Table 4-5. Instrument Error Byte (FW rev 1.12, logged, displayed as E2:xx during errors)
Bit Description
E/W Conditions
Front Panel Message
0
Bench Comm
Error
E
Bench is not responding
BENCH COM ERROR
1
n/a
-
Not implemented
2
DAC Warning
W
DAC is unresponsive
DAC WARNING
3
RTC Error
E
RTC is malfunctioning
RTC ERROR
4
EEPROM Warning W
EEPROM was initialized
EEPROM WARNING
5
AFE Warning
W
A-to-D system on motherboard is
unresponsive
AFE WARNING
6
Pump Warning
W
Pressure drop after starting pump is
too small
PUMP WARNING
7
n/a
-
TECHNICAL ASSISTANCE
4.3.
Contact Teledyne API Technical Support for any assistance:
Phone: +1 858-657-9800, Toll-free: 800-324-5190