Operation
Teledyne API – Model 430 Compact Ozone Analyzer
20
08311C DCN7342
DATA
3.2.4.
The Data menu allows configuration of the Analog Output and operation of the internal
data log and the SD Card. Note that the menus use “NV” (non-volatile) to refer to data
in the internal memory as distinguished from data on an SD card).
Table 3-3. Data Menu Description
Menu Item
Description
Analog Out
Analog Out Range
and
Analog Out Offset
are used to configure the
analog output for 0-5 V, which correlates to the 0-to-maximum
concentration range, by adjusting the voltage values through their
respective Edit menus. The voltage output signal is automatically scaled
according to the settings.
Analog Out Step Test
ramps the analog output signal from 0 to 5 V at
1.25 V increments to test the analog output. The analog output signal can
be verified by measuring the voltage with a multimeter.
Data Log
Data Log Send
sends the data stored on the internal memory through the
serial output.
Data Log # Days>Edit
sets the number of days’ worth of data that will be
sent under the Data Log Send command.
Clear NV
Log
Clear NV Log
clears the data from the non-volatile internal memory log.
(Does not impact any data on an SD card).
SD Card
NV Log to SD Card
sends the data stored on the non-volatile (NV) internal
memory to the SD card.
DO NOT REMOVE THE SD CARD OR POWER OFF THE
INSTRUMENT WITHOUT FIRST RUNNING A SOFTWARE EJECT VIA
THE DATA>
EJECT SD CARD
MENU. Improper ejection of the SD card
may result in corruption or loss of data.
Eject SD Card
performs a software eject, which is required
prior
to
removing the SD card or powering off the instrument.