Teledyne API – Model 430 Compact Ozone Analyzer
Getting Started
08311 DCN7342
13
PNEUMATIC CONNECTIONS
2.2.
SAMPLE IN
Connect ¼” OD tubing (PTFE recommended) to the SAMPLE IN
fitting.
EXHAUST
Connect ¼” OD tubing (PTFE recommended) to the EXHAUST
fitting; ensure maximum 10 m length and the exhaust gases are
properly ventilated.
INTERNAL PUMP
2.2.1.
The instrument is equipped with an internal pump rated at 10,000 hours and requires no
additional connections; refer to Section 4.1 for the maintenance schedule and Section 6
for the part number.
ELECTRICAL CONNECTIONS
2.3.
Note
To maintain compliance with EMC standards, it is required that the cable
length be no greater than 3 meters for all I/O connections, which include
Power, Analog In, Analog Out, RS-232, and RS-485.
The Model 430 rear panel (Figure 1-2) has connectors for power, for serial
communications, and for data collection and reporting.
POWER
2.3.1.
Connect the power cord provided with the instrument to the 12 V DC power port. (If
using the external battery option, insert the battery connector into the power port). The
polarity of the power port is center positive as shown below:
Figure 2-1. Power Port Polarity
CAUTION!
Ensure that the power source complies with the power
rating for this instrument.
Always maintain easy accessibility for quickly
disconnecting power if needed.