4250 Flow Meter
Section 2 Programming
2-30
You must have an Isco
674 Rain Gauge
(or approved equiv-
alent) connected to the flow meter through the
Rain Gauge
Port
to sense rainfall. MM = millimeters. The rain gauge is
factory-calibrated. See Section 4 for more information about the
rain gauge. If you are not using a rain gauge, you would select
NOT MEASURED for this step.
pH measurement determines the relative acidity or alkalinity of
a solution. You must have an Isco
pH Probe
(or approved equiv-
alent) connected to the flow meter through the parameter port to
sense pH. pH measurements range from 0 to 14 pH units, with
solutions below 7 considered acidic and solutions above 7,
alkaline. Pure water has a pH of 7.
These devices require periodic recalibration for accurate sensing
of pH. See Section 4 for more information about the pH probe. If
you are not using a pH probe you would select NOT MEASURED
for this step.
• If you choose NOT MEASURED for pH, no more refer-
ences to pH will appear for the rest of the program.
• If you select pH, you will be unable to measure D. O. and
references to D.O. will not appear on the rest of the
program.
• If your situation requires the measurement of both
parameters (pH and D.O.) at the same time, or if you
also need to measure conductivity, you should use the
YSI 600 Multi-Parameter Sonde.
If you select NOT MEASURED and press
Enter
, the following
will appear:
Measurement of dissolved oxygen is conducted in studies of
water quality in lakes and streams. Some dissolved oxygen is
necessary for the survival of aquatic life in these waters.
You must have an Isco
Dissolved Oxygen Probe
(or approved
equivalent) to sense dissolved oxygen. The probe attaches to the
Parameter Port
. These devices require periodic calibration for
accurate sensing. See Section 4 for more information about the
dissolved oxygen probe. PPM = parts per million; MG/L = milli-
grams per liter.
If you are not using the D. O. probe, you would select NOT MEA-
SURED for this step.
Selection of PPM or MG/L will keep references to pH from
showing up on subsequent menus.
RAIN GAUGE
• INCHES • • MM • • NOT MEASURED •
pH UNITS OF MEASURE
• pH • • NOT MEASURED •
D. O. UNITS
• MG/L • • PPM • • NOT MEASURED •
Содержание 4250
Страница 2: ......
Страница 4: ......
Страница 12: ...4250 Flow Meter Table of Contents x ...
Страница 28: ...4250 Flow Meter Section 1 Introduction 1 16 ...
Страница 90: ...4250 Flow Meter Section 2 Programming 2 62 ...
Страница 140: ...4250 Flow Meter Section 4 Options and Accessories 4 30 ...
Страница 160: ...4250 Flow Meter Appendix A Replacement Parts and Accessories A 2 ...
Страница 161: ...4250 Flow Meter Appendix A Replacement Parts and Accessories A 3 ...
Страница 166: ...4250 Flow Meter Appendix A Replacement Parts and Accessories A 8 ...
Страница 184: ...4250 Flow Meter Appendix C General Safety Procedures C 10 ...
Страница 191: ...4250 Flow Meter Appendix D Material Safety Data Sheets D 7 ...
Страница 192: ...4250 Flow Meter Appendix D Material Safety Data Sheets D 8 ...
Страница 196: ......
Страница 198: ......
Страница 200: ......
Страница 202: ...Warranty ...
Страница 204: ......