4250 Flow Meter
Section 4 Options and Accessories
4-26
Figure 4-10 D.O. Parameter Module
You must use the Isco 270 D.O. Module box between the probe
and flow meter; this extends the distance between the probe and
the flow meter to 1,000 feet.
4.11.6 Calibrating the D.O.
Probe with a Flow
Meter
Prepare the probe as described above if this has not already been
done (fill the probe cavity with electrolyte and seat the mem-
brane).
Note
You must use the Isco Temperature Probe with the D.O. Probe
to provide temperature compensation.
Wrap both the D.O. Probe and Temperature Probe in a damp
cloth. Wait ten minutes for it to stabilize, then proceed.
Go to
step 1
on the flow meter. Select PROGRAM, then step
through the units of measure with
Enter
until you reach the
menu for pH measurement. Select NOT MEASURED. Press
Enter
. Then D.O. UNITS will appear. Select either MG/L or
PPM. Press
Enter
. Exit the program and re-enter going to
step
3
.
You must select measurement of D.O. in
step 1
or the D.O. menu
will not appear in
step 3
. If you cannot get the D.O. menu to
appear in
step 3
, recheck your selections in
step 1
. Remember
that you must select NOT MEASURED for pH in
step 1
or D.O.
will not appear on the menu for the rest of the program.
Select D.O. Press
Enter
. This display will appear:
SELECT PARAMETER TO ADJUST
NONE LEVEL D.O.
D.O. CALIBRATION
D.O STANDARD, ABS BAROMETRIC PRESSURE,
Содержание 4250
Страница 2: ......
Страница 4: ......
Страница 12: ...4250 Flow Meter Table of Contents x ...
Страница 28: ...4250 Flow Meter Section 1 Introduction 1 16 ...
Страница 90: ...4250 Flow Meter Section 2 Programming 2 62 ...
Страница 140: ...4250 Flow Meter Section 4 Options and Accessories 4 30 ...
Страница 160: ...4250 Flow Meter Appendix A Replacement Parts and Accessories A 2 ...
Страница 161: ...4250 Flow Meter Appendix A Replacement Parts and Accessories A 3 ...
Страница 166: ...4250 Flow Meter Appendix A Replacement Parts and Accessories A 8 ...
Страница 184: ...4250 Flow Meter Appendix C General Safety Procedures C 10 ...
Страница 191: ...4250 Flow Meter Appendix D Material Safety Data Sheets D 7 ...
Страница 192: ...4250 Flow Meter Appendix D Material Safety Data Sheets D 8 ...
Страница 196: ......
Страница 198: ......
Страница 200: ......
Страница 202: ...Warranty ...
Страница 204: ......