4250 Flow Meter
Section 2 Programming
2-16
tative samples. However, if your stream shows continuous flow
and the sensor detects an error, you would probably select NO
(use previous reading).
If you select DO/PH READING INTERVAL, the following will
appear:
Again, this selection is offered as a means of saving battery
power. After selecting the appropriate parameter reading
interval, press
Enter
. The
Setup
menu will reappear. This time,
select YSI 600 READING INTERVAL from the menu. The fol-
lowing display will appear:
After selecting the appropriate parameter reading interval, press
Enter
. The
Setup
menu will reappear. This time, select STATUS
from the
Setup
menu. Press
Enter
. The following will appear:
HW REV refers to the hardware revision number.
SW REV refers to the software revision number.
ID is an identification number for the flow meter.
If you press
Enter
again, you will see the following display:
This value indicates the functioning of the area-velocity sensor. (The
flow meter goes into the continuous reading mode whenever this
screen is displayed.) This display exists primarily as a diagnostic
tool for telephone troubleshooting. There is no “magic” number you
should expect to see. If the flow meter is displaying valid level/flow
readings, you do not need to worry much about the numbers. The
numbers become important when you
cannot
get valid level/flow
readings. You can apply the following explanations to the values
displayed.
DO/PH READING INTERVAL
• CONTINUOUS • • 15 SEC • • 30 SEC • • 1 MIN • • 2 MIN • • 5 MIN •
YSI 600 READING INTERVAL
•CONTINUOUS••15 SEC••30 SEC••1 MIN••2 MIN••5 MIN•
MODEL 4250 HW REV:
XXXXXX
SW REV X.XX
ID
XXXXXXXXXXX
SIGNAL STRENGTH
X
%
SPECTRUM STRENGTH
X
%
Signal
Strength
Spectrum
Strength
Explanation
0%
0%
Unit is approximating velocity.
(Appears when level is below
minimum velocity depth.)
0%
ERR
Not enough reflective particles.
ERR
ERR
Not working—call factory.
Содержание 4250
Страница 2: ......
Страница 4: ......
Страница 12: ...4250 Flow Meter Table of Contents x ...
Страница 28: ...4250 Flow Meter Section 1 Introduction 1 16 ...
Страница 90: ...4250 Flow Meter Section 2 Programming 2 62 ...
Страница 140: ...4250 Flow Meter Section 4 Options and Accessories 4 30 ...
Страница 160: ...4250 Flow Meter Appendix A Replacement Parts and Accessories A 2 ...
Страница 161: ...4250 Flow Meter Appendix A Replacement Parts and Accessories A 3 ...
Страница 166: ...4250 Flow Meter Appendix A Replacement Parts and Accessories A 8 ...
Страница 184: ...4250 Flow Meter Appendix C General Safety Procedures C 10 ...
Страница 191: ...4250 Flow Meter Appendix D Material Safety Data Sheets D 7 ...
Страница 192: ...4250 Flow Meter Appendix D Material Safety Data Sheets D 8 ...
Страница 196: ......
Страница 198: ......
Страница 200: ......
Страница 202: ...Warranty ...
Страница 204: ......