Installation Series
4060
32
Teledyne
Analytical
Instruments
40
0.4
10.4
50
0.5
12.0
60
0.6
13.6
70
0.7
15.2
80
0.8
16.8
90
0.9
18.4
100
1.0
20.0
Table 3-3: Analog Range ID Output—Example
Range
Voltage Signal
Current Signal
Application
Output (VDC)
Output (mA DC)
Range 1
0.20
7.2 0–100 ppb Benzene
Range 2
0.50
12 0–1 ppm Benzene
Range 3
0.80
16.8 0–10 ppm Benzene
3.3.2.5
A
LARM
R
ELAYS
The nine alarm-circuit connector pins connect to the internal alarm
relay contacts. Each set of three pins provides one set of Form C relay
contacts. Each relay has both normally open and normally closed contact
connections. The contact connections are shown in Table 3-4. They are
capable of switching up to 3 amperes at 250 VAC into a resistive load.
The connectors are:
Threshold Alarm 1
:
Can be configured as high (actuates when
concentration is above threshold), or low (actuates
when concentration is below threshold).
Can be configured as failsafe or non-failsafe
Can be configured as non-latching
Can be configured out (defeated).