Series 4060
Installation
Teledyne
Analytical
Instruments
35
2.
When the CRC closes, remove the span command.
When CRC opens again, the span calibration is done, and the sample
is being analyzed.
3.3.2.7
R
ANGE
ID
R
ELAYS
There are three dedicated Range ID relay contacts. They are
assigned to relays in ascending order—Low range is assigned to Range
1 ID, Medium range is assigned to Range 2 ID, and High range is
assigned to Range 3 ID. Table 3-6 lists the pin connections. Contacts
are normally open, and they close when 4060 switches to that particular
range.
Table 3-6: Range ID Relay Connections
Pin
Function
21
Range 1 ID Contact
38
Range 1 ID Contact
22
Range 2 ID Contact
39
Range 2 ID Contact
19
Range 3 ID Contact
18
Range 3 ID Contact
34
Not
Used
35
Not
Used
3.3.2.8
N
ETWORK
I/O
A serial digital input/output for local network protocol. At this
printing, this port is not yet functional. It is to be used in future options
to the instrument. Pins 13 (+) and 29 (–).
3.3.2.9
P
IN
O
UT
T
ABLE
The following table summarizes all the outputs/inputs available in
the 50 pin D-Sub connector on the back panel of the analyzer.