Oxygen Analyzer
Part II: Analysis Unit
Part II: 4-3
Table 4-1: Self Test Failure Codes
Power
0
OK
1
5 V Failure
2
15 V Failure
3
Both Failed
Analog
0
OK
1
DAC A (0–1 V Concentration)
2
DAC B (0–1 V Range ID)
3
Both Failed
Preamp
0
OK
1
Zero too high
2
Amplifier output doesn't match test input
3
Both Failed
The results screen alternates for a time with:
Press Any Key
To Continue...
Then the analyzer returns to the initial System screen.
4.5
Cell Failure Checks
Cell failure is covered in detail in Part I: Control Units, section
4.4.1.3, Cell Failure. Cell replacement is covered Part II: Analysis Units
chapter 5, Maintenance.
When the sensor in the 3010P begins to fail, the analyzer will usually
require more and more frequent calibration. If recalibration raises the
readings temporarily only, suspect the cell.
You can check the output of the cell itself by scrolling the MAIN
MENU TO
SENSOR.
When you ENTER the function, the sensor report screen appears.
RAW CELL OUTPUT
###
µµµµµ
A
Содержание 3010PA
Страница 17: ...1 6 Part I 1 Introduction Model 3010PA ...
Страница 19: ...2 2 Part I 2 Operational Theory Model 3010PA Figure 2 1 Block Diagram of the 3010PA CU Electronics ...
Страница 21: ...2 4 Part I 2 Operational Theory Model 3010PA ...
Страница 57: ...iv Part II Model 3010P Oxygen Analyzer ...
Страница 61: ...1 4 Part II 1 Introduction 1 Introduction 1 Introduction 1 Introduction 1 Introduction Model 3010P ...
Страница 69: ...2 8 Part II 2 Operational Theory Model 3010P ...
Страница 77: ...3 Installation Model 3010P 3 8 Part II ...
Страница 101: ...A 12 Appendix Appendix Models 3010PA ...