Oxygen Analyzer
Part II: Analysis Unit
Part II: 3-7
Figure 3-6: FET Series Resistance
3.5
Installing the Micro-Fuel Cell
The Micro-Fuel Cell is not installed in the cell block when the
instrument is shipped. It must be installed during initial installation.
Once it is expended, or if the cell is exposed to air for too long, the
Micro-Fuel Cell will need to be replaced. The cell could also require replace-
ment if the instrument has been idle for too long.
When the micro-Fuel Cell needs to be installed or replaced, follow the
procedures in chapter 5, Maintenance, for removing and installing cells.
3.6
Testing the System
After The Control Unit and the Analysis Unit are both installed and
interconnected, and the system gas and electrical connections are complete,
the system is ready to test. Before plugging either of the units into their
respective power sources:
•
Check the integrity and accuracy of the gas connections. Make
sure there are no leaks.
•
Check the integrity and accuracy of the electrical connections.
Make sure there are no exposed conductors
•
Check that sample pressure is between 3 and 40 psig, according
to the requirements of your process.
Power up the system, and test it as follows:
1. Repeat the Self-Diagnostic Test as described in Part I, chapter 4,
section 4.3.5.
Содержание 3010PA
Страница 17: ...1 6 Part I 1 Introduction Model 3010PA ...
Страница 19: ...2 2 Part I 2 Operational Theory Model 3010PA Figure 2 1 Block Diagram of the 3010PA CU Electronics ...
Страница 21: ...2 4 Part I 2 Operational Theory Model 3010PA ...
Страница 57: ...iv Part II Model 3010P Oxygen Analyzer ...
Страница 61: ...1 4 Part II 1 Introduction 1 Introduction 1 Introduction 1 Introduction 1 Introduction Model 3010P ...
Страница 69: ...2 8 Part II 2 Operational Theory Model 3010P ...
Страница 77: ...3 Installation Model 3010P 3 8 Part II ...
Страница 101: ...A 12 Appendix Appendix Models 3010PA ...