Appendix
3000ZA
Teledyne Analytical Instruments
To improve the system response, a by-pass can be added to increase
the sample flow rate to the analyzer by a factor of ten. A by-pass provides
a sample flow path around the analyzer of 2 - 18 SCFH. typically.
CALIBRATION GAS:
3000 series analyzer requirements for units with Auto-Cal options
are as follows:
The customer must supply control valves (or restrictors) for any SPAN
or ZERO gas source which is attached to the Auto-Cal ports. The valve
should be adjusted to the same flow rate as the sample gas. When restrictors
are used, the gas pressure must be adjusted to achieve the proper flow rate.
OPERATION WITHOUT A RESTRICTOR DEVICE:
Operation without a restrictor device is not recommend as
mentioned above. A 3000TA without any flow restrictor device was
tested on 11-19-97. This results in a flow rate of 2.4 SLPM @ 1 PSIG.
This is a cv of 0.023 for the standard sample system.
REFERENCE: FLOW_1.XLS & FLOW_2.XLS for information on
flow rates at various pressures.
TAI PART NUMBERS
Restrictor kit:
A68729
Union (SS)
U11
LP Restrictor
R2323 (low pressure / vac. service)
Std. Restrictor
R2324 Blue dot
Nut
N73
Ferrule
F73
Ferrule
F74
Both ferrules are required
CONVERSIONS:
1 PSI
=
2.04 inches of mercury (in. Hg.)
Содержание 3000ZA
Страница 24: ...Operational Theory 3000ZA Teledyne Analytical Instruments 12 Figure 2 3 Component Location...
Страница 25: ...Trace Oxygen Analyzer Operational Theory Teledyne Analytical Instruments 13 Figure 2 4 Electronic Block Diagram...
Страница 63: ...Trace Oxygen Analyzer Operation Teledyne Analytical Instruments 51...
Страница 77: ...Trace Oxygen Analyzer Appendix Teledyne Analytical Instruments 1 SCFH 0 476 SLPM...
Страница 81: ...Trace Oxygen Analyzer Index Teledyne Analytical Instruments...