Trace Oxygen Analyzer
Operation
Teledyne Analytical Instruments
49
as established by the operator for that particular range, the
output saturates at 1 VDC(or 20 mA). However, the digital
readout and the RS-232 output continue to read the true
value of the oxygen concentration regardless of the analog
output range.
4.7 The Analyze Function
Normally, all of the functions automatically switch back to the
Analyze function when they have completed their assigned operations.
Pressing the
Escape
button in many cases also switches the analyzer
back to the Analyze function. Alternatively, you can press the
Analyze
button at any time to return to analyzing your sample.
4.8 Signal Output
The standard Model 3000ZA Trace Oxygen Analyzer is equipped
with two 0–1 VDC analog output terminals accessible on the back panel
(one concentration and one range ID), and two isolated 4–20 mA DC
current outputs (one concentration and one range ID).
See
Rear Panel
in Chapter 3,
Installation
for connection details.
The signal output for concentration is linear over the currently
selected analysis range. For example, if the analyzer is set on range that
was defined as 0–100 ppm O
2
, then the output would be:
Table 4-1: Output Signals
Voltage Signal
Current Signal
ppm O2
Output (VDC)
Output (mA DC)
0
0.0
4.0
10
0.1
5.6
20
0.2
7.2
30
0.3
8.8
40
0.4
10.4
50
0.5
12.0
60
0.6
13.6
70
0.7
15.2
Содержание 3000ZA
Страница 24: ...Operational Theory 3000ZA Teledyne Analytical Instruments 12 Figure 2 3 Component Location...
Страница 25: ...Trace Oxygen Analyzer Operational Theory Teledyne Analytical Instruments 13 Figure 2 4 Electronic Block Diagram...
Страница 63: ...Trace Oxygen Analyzer Operation Teledyne Analytical Instruments 51...
Страница 77: ...Trace Oxygen Analyzer Appendix Teledyne Analytical Instruments 1 SCFH 0 476 SLPM...
Страница 81: ...Trace Oxygen Analyzer Index Teledyne Analytical Instruments...