Web Implementation
3000TA-XL-EU
Teledyne Analytical Instruments
68
5.3 Set the Alarms Remotely
To set the concentration alarms, Alarm 1 and Alarm 2, click on the
Alarm
button. The setup menu for both alarms will appear on the left
side of the web page. After a few seconds, the page will be updated with
current settings on the alarms. To find out what each function does, refer
to the alarm section in Chapter 4. Only one alarm can be set each time.
Adjust all the parameters on the alarm of interest, then click the
Set
button for the corresponding alarm to update control unit. It takes about
five seconds for the update to occur. Do not attempt adjustments while
‗
Getting Data
‘ is displayed in the message window. Repeat the process
for adjusting the second alarm.
Figure 5-4: Set Alarms Window