Operation
3000TA-XL-EU
Teledyne Analytical Instruments
56
For non-failsafe operation, the relay is energized in an
alarm condition.
4.
Are either of the alarms to be latching?
In latching mode, once the alarm or alarms trigger,
they will remain in the alarm mode even if process
conditions revert back to non-alarm conditions. This
mode requires an alarm to be recognized before it can
be reset.
In the non-latching mode, the alarm status will
terminate when process conditions revert to non-
alarm conditions.
5.
Are either of the alarms to be defeated?
The defeat alarm mode is incorporated into the alarm
circuit so that maintenance can be performed under
conditions which would normally activate the alarms.
The defeat function can also be used to reset a latched
alarm. (See procedures, below.)
If you are using password protection, you will need to enter your
password to access the alarm functions. Follow the instructions in
Section 4.3.3 to enter your password. Once you have clearance to
proceed, enter the
Alarm
function.
Press the
Alarm
button on the front panel to enter the Alarm
function. Make sure that
AL–1
is blinking.
AL—1 AL—2
Choose Alarm
Set up alarm 1 by moving the blinking over to AL–1 using the ◄►
arrow keys. Then press
Enter
to move to the next screen.
AL—1 1000 ppm HI
Dft—N Fs—N Ltch—N
Five parameters can be changed on this screen:
Value of the alarm setpoint,
AL–1 ####
ppm (oxygen);
this value can be set from 0.00 to 10,000.00 ppm
Out-of-range direction,
HI
or
LO
Defeated?
Dft–Y/N
(Yes/No)
Содержание 3000TA-XL-EU
Страница 3: ...Trace Oxygen Analyzer Teledyne Analytical Instruments iii...
Страница 16: ......
Страница 29: ...Trace Oxygen Analyzer Operational Theory Teledyne Analytical Instruments 13 Figure 2 4 Piping Layout...
Страница 34: ...Operational Theory 3000TA XL EU Teledyne Analytical Instruments 18...
Страница 52: ...Installation 3000TA XL EU Teledyne Analytical Instruments 36...
Страница 82: ...Web Implementation 3000TA XL EU Teledyne Analytical Instruments 66 Figure 5 2 Span Calibration Window...
Страница 88: ...Web Implementation 3000TA XL EU Teledyne Analytical Instruments 72...
Страница 92: ...Maintenance 3000TA XL EU Teledyne Analytical Instruments 76 Figure 6 1 Removing the Micro Fuel...