Installation
3000TA-XL-EU
Teledyne Analytical Instruments
28
1 ID, Medium range is assigned to Range 2 ID, and High range is
assigned to Range 3 ID. The fourth range is reserved for the Air Cal
Range (25%). Table 3-4 lists the pin connections.
Table 3-4: Range ID Relay Connections
Pin
Function
21
Range 1 ID Contact
38
Range 1 ID Contact
22
Range 2 ID Contact
39
Range 2 ID Contact
19
Range 3 ID Contact
18
Range 3 ID Contact
34
Range 4 ID Contact (Air Cal)
35
Range 4 ID Contact (Air Cal)
3.3.2.2.7 Network I/O
A serial digital input/output for local network protocol. At this
printing, this port is not yet functional. It is to be used for future options
to the instrument. Pins 13 (+) and 29 (–).
3.3.2.2.8 Remote Valve Connections
The 3000TA-XL-EU is a single-chassis instrument, which has no
Remote Valve Unit. Instead, the Remote Valve connections are used as
a method for directly controlling external sample/zero/span gas valves.
See Figure 3-5.
Содержание 3000TA-XL-EU
Страница 3: ...Trace Oxygen Analyzer Teledyne Analytical Instruments iii...
Страница 16: ......
Страница 29: ...Trace Oxygen Analyzer Operational Theory Teledyne Analytical Instruments 13 Figure 2 4 Piping Layout...
Страница 34: ...Operational Theory 3000TA XL EU Teledyne Analytical Instruments 18...
Страница 52: ...Installation 3000TA XL EU Teledyne Analytical Instruments 36...
Страница 82: ...Web Implementation 3000TA XL EU Teledyne Analytical Instruments 66 Figure 5 2 Span Calibration Window...
Страница 88: ...Web Implementation 3000TA XL EU Teledyne Analytical Instruments 72...
Страница 92: ...Maintenance 3000TA XL EU Teledyne Analytical Instruments 76 Figure 6 1 Removing the Micro Fuel...
Страница 108: ...Appendix 3000TA XL EU Teledyne Analytical Instruments 92...