2105 Interface Module
Section 2 Installation and Operation

2-8

2.4 Antenna (2105Ci Only)

The 2105Ci uses a magnetic mount antenna and Serial Over
Internet Protocol (SOIP) communication to connect with other
units. (The module is also capable of CSD communication, but
this service is being phased out. See Section 2.5.1 for complete
information.) Phone service with a static IP address, rather than
a land line, provides efficient communication.

Note

The 2105Ci only works in a CDMA cellular service area with

1xRTT. See your local CDMA service provider for availability.

As you face the unit, the 2105Ci antenna connector socket is

located on the right hand side of the module (Figure 2-9). Remove
the connector cap from the socket and attach it to the cap on the
end of the antenna cable. To connect the antenna, press down on
the silver metal latch on the side of the module and push the
antenna’s connector into the socket.

CAUTION

Only use this antenna with 2100 Series Ci cellular modem
modules. Do not use this antenna with non-Ci equipment.

Figure 2-9

Antenna connected to 2105Ci

Note

For the transmitter to comply with FCC Maximum Permissible
Exposure (MPE) regulations, the antenna must be located a
minimum of 30 centimeters (12 inches) from the human body.

Note

When any communication connector is not in use, it should

always be capped. The cap will seal the connector to prevent
corrosion, prevent moisture from entering the unit, and improve
communications.

Содержание 2105

Страница 1: ...2105 2105Ci Interface Module Installation and Operation Guide Part 69 2003 635 of Assembly 60 2004 635 Copyright 2011 All rights reserved Teledyne Isco Revision B October 15 2012 ...

Страница 2: ......

Страница 3: ...rization Number specified Be sure to include a note describing the malfunction This will aid in the prompt repair and return of the equipment Teledyne Isco welcomes suggestions that would improve the information presented in this manual or enhance the operation of the equipment itself Teledyne Isco is continually improving its products and reserves the right to change product specifications replac...

Страница 4: ......

Страница 5: ...anière autre que celles qui sont specifiées dans ce manuel la protection fournie de l instrument peut être affaiblie cela augmentera votre risque de blessure Hazard Severity Levels This manual applies Hazard Severity Levels to the safety alerts These three levels are described in the sample alerts below CAUTION Cautions identify a potential hazard which if not avoided may result in minor or modera...

Страница 6: ...s de sécurité Ce symbole signale l existence d instructions importantes relatives au produit dans ce manuel Ce symbole signale la présence d un danger d électocution Warnungen und Vorsichtshinweise Das Ausrufezeichen in Dreieck ist ein Warnzeichen das Sie darauf aufmerksam macht daß wichtige Anleitungen zu diesem Handbuch gehören Der gepfeilte Blitz im Dreieck ist ein Warnzeichen das Sei vor gefäh...

Страница 7: ...Connecting to Flowlink 2 9 2 5 1 Modem Site Connection 2 12 2 6 Setting Up Text Messaging 2105Ci 2 16 2 7 Power Conservation 2105Ci 2 17 2 7 1 Equation 2 19 2 8 Pushed Data Capability 2105Ci 2 20 2 9 SDI 12 Input 2 21 2 10 Rain Gauge Interface 2 22 2 11 Sampler Interface 2 23 2 12 Sampler Pass Thru Mode 2 24 2 13 Pulse Doppler Devices 2 28 2 14 Modbus Input 2 29 2 14 1 4200 and 6700 Series Measure...

Страница 8: ...i serial label 2 1 2 2 Identifying the 2150 voltage specification 2 2 2 3 Identifying the voltage specification on the DB9 cable 2 3 2 4 Unlocking the Latch 2 4 2 5 Locking the Latch 2 5 2 6 Loading the 2191 Battery module 2 6 2 7 Unlock latch and stow the cap 2 7 2 8 Aligning the modules 2 7 2 9 Antenna connected to 2105Ci 2 8 2 10 Flowlink connect screen version 5 10 616 2 9 2 11 Site resolution...

Страница 9: ...e connector 2 38 3 1 Configuration example direct connection shown 3 3 List of Tables 1 1 2105 Interface Module Components Top View 1 2 1 2 2105 Interface Module Components Bottom View 1 3 1 3 Battery Module Components Top View 1 4 1 4 Battery Module Components Bottom View 1 5 1 5 Technical Specifications 2105 Interface Module 1 6 1 6 Technical Specifications 2191 Battery Module 1 7 2 1 Voltage Sp...

Страница 10: ...2105 Interface Module Table of Contents viii ...

Страница 11: ...version of the Isco Model 677 logging rain gauge Isco rain gauge and 2105 module combination your system includes the rain gauge interface cable part 69 2004 581 The 2105 can be located anywhere within a stack of up to three other 2100 Series networked modules using the same locking mechanism that connects the 2100 Series modules to each other The 2105 is compatible with Isco s 2150 Area Velocity ...

Страница 12: ...t in use to protect the connector from moisture damage When the connector is in use store the connector cap on the cap holder 3 Cap Holder Stores the connector cap when the communication connector is in use 4 Instrument Interface Cable Connector Used to connect the module to compatible equipment for data logging and sampler control 5 Connector Plug Insert into the interface connector when the conn...

Страница 13: ...in the stack and is used to transfer data and or receive power 2 Connector Plug Insert into the communication connector when not in use to protect the connector from moisture damage When the connector is in use store the connector cap in the cap holder 3 Plug Holder Stores the connector Plug when the communication connector is in use 4 Desiccant Cartridge and Hydrophobic Filter Prevents moisture f...

Страница 14: ...tack transfers power and data 2 Connector Cap Stowed on Cap Holder Insert into the communication connector when not in use to protect the connector from moisture damage When the connector is in use store the connector cap on the cap holder 3 Lantern Battery Alkaline shown 6V alkaline or rechargeable lead acid battery quantity of 2 4 Door Two circular doors contain the desiccant bags hold the batte...

Страница 15: ...on Connector Connects the modules in the stack transfers power and data 2 Connector Plug Insert into the communication connector when not in use to protect the con nector from moisture damage When the connector is in use store the con nector cap in the cap holder 3 Plug Holder Stores the connector plug when the communication connector is in use 4 Latch Push in to lock the module in a stack ...

Страница 16: ...ation Speeds Supported 9600 19200 and 38400 bps Optional Cellular Communication Serial Over IP or CSD 2105CI Data Storage 798 Kb non volatile flash Storage Mode Rollover 5 bytes per reading Storage Interval 15 or 30 seconds 1 2 5 15 or 30 minutes or 1 2 4 12 or 24 hours Storage rate variable based on measured parameters Data Types Flow Rate Level Rainfall Conductivity Dissolved Oxygen Temperature ...

Страница 17: ... warn you of unsafe practices or conditions that may cause property damage WARNING Warnings indicate potentially hazardous conditions If you do not avoid these risks they could cause you death or serious injury Table 1 6 Technical Specifications 2191 Battery Module Dimensions Width 9 6 inches 24 4 cm Height 6 0 inches 15 2 cm Depth 7 6 inches 19 3 cm Weight without batteries 3 2 lbs 1 4 kg Materia...

Страница 18: ...nt for assis tance Teledyne Isco Technical Service Department P O Box 82531 Lincoln NE 68501 866 298 6174 or 402 464 0231 FAX 402 465 3001 e mail IscoService teledyne com Simple difficulties can often be diagnosed over the phone If it is necessary to return the equipment to the factory for service please follow the shipping instructions provided by the Technical Service Department including the us...

Страница 19: ...y be other packing lists for various sub components It is recommended that you retain the shipping cartons as they can be used to ship the unit in the event that it is necessary to transport the system Please complete the registration card and return it to Teledyne Isco 2 1 1 2105Ci Label Information If you have a 2105Ci you should record the IP address and phone number printed on the label found ...

Страница 20: ...module crown connector on the earlier RS 232 DB9 digital communication cable part 60 2004 046 is unmarked The con nector on the 26 volt cable has a serial tag specifying the higher voltage refer to Figure 2 3 Table 2 1 Voltage Specifications for 2100 System Components Module or Cable Earlier Voltage Range Current Voltage Range Updated 2150 7 16 6 VDC 7 26 VDC March 2005 2110 7 16 6 VDC N A 2101 21...

Страница 21: ...s digesters and storage tanks These spaces may become hazardous environments that can prove fatal for those unpre pared These spaces are governed by OSHA 1910 146 and require a permit before entering WARNING Avoid hazardous practices If you use these instruments in any way not specified in this manual the protection provided by the instruments may be impaired this will increase your risk of injury...

Страница 22: ...ating the latches The latch is normally locke but you must unlock the latch to place the module on top of another module in a stack The latch is unlocked by pushing in the latch release on the connector side of the module To lock the latch push in the latch on the desiccant side of the module CAUTION The latch can be damaged by applying too much force Never press on both sides at the same time Do ...

Страница 23: ... capped The cap will seal the con nector to prevent corrosion prevent moisture from entering the unit and improve communications When a communication connector is in use store the cap on the holder next to the connector The communication connector will be sealed by its mating connector CAUTION Caps PUSH ON and PULL OFF Do not rotate the caps to remove them from the connectors Note For modules to c...

Страница 24: ...tteries follow the instructions below Figure 2 6 Loading the 2191 Battery module Note The batteries should always be replaced as a pair Never mix old and new batteries 1 Load the lantern battery into the carrier 2 Insert the carrier and battery into the module Note the position of the carrier s connector it must be aligned toward the center of the module 3 Align the marks indicated and insert the ...

Страница 25: ...mmunication connector It should be clean and dry Damaged O rings must be replaced 3 If you are using the metal carrying handle insert it between the top two modules with the handle turned toward the rear of the stack opposite the yellow labels 4 Unlock the 2105 s latch by pressing in on the latch release 5 Underneath the 2105 remove the cap from the lower com munication connector and stow it in th...

Страница 26: ...ocated on the right hand side of the module Figure 2 9 Remove the connector cap from the socket and attach it to the cap on the end of the antenna cable To connect the antenna press down on the silver metal latch on the side of the module and push the antenna s connector into the socket CAUTION Only use this antenna with 2100 Series Ci cellular modem modules Do not use this antenna with non Ci equ...

Страница 27: ... 2004 046 or USB port connect cable part 60 2004 507 Open Flowlink and go to the Connect screen Figure 2 10 by either selecting it from the pull down menu or clicking on the Quick Connect icon Make sure the connection Type is Direct and click on the 2100 Instrument icon to connect Upon initial connection Flowlink creates a site file and adds it to the database If the system detects the addition of...

Страница 28: ...re 2 11 Site resolution screen To add the new module to an existing site select the appropriate site and click OK To create a new site select Create a New Site Click in the name field enter the name for the site and click OK Upon connection the Site Info tab will appear ...

Страница 29: ...2105 Interface Module Section 2 Installation and Operation 2 11 Figure 2 12 Site Information screen ...

Страница 30: ... the modem s operation on the Modem screen The 2105Ci s digital cellular modem provides a choice of two dif ferent connection methods Serial Over IP IP or Circuit Switched Data CSD The default Contact Connection Mode is IP this mode of communication is much faster and does not require analog infrastructure Note CSD communication is being phased out by the service pro vider Verizon As of July 2012 ...

Страница 31: ...ion 2 Installation and Operation 2 13 The APN settings are for GSM modems only and do not apply to this equipment Figure 2 14 Modem screen configuring the cellular modem Select IP recommended or CSD TCP port number default 1700 ...

Страница 32: ...rial over IP connection select the TCP radio button on the Devices tab In the TCP address field to the left of the colon enter the 2105Ci modem s IP address found on the serial tag on the back of the module The default TCP port is 1700 but can be edited if necessary Click Apply to save your set tings Figure 2 15 Devices screen configuring the modem for Serial Over IP communication Select TCP radio...

Страница 33: ... for this site only click the arrow to view a pull down list containing all modems recognized by your Windows operating system Lastly enter the phone number of the 2105Ci s modem found on the serial tag on the back of the module Click Apply to save your settings Figure 2 16 Devices screen configuring the modem for CSD communication Select Modem radio button Enter cell phone number Select Windows m...

Страница 34: ...d alarm condition occurs To program the 2105Ci for text messaging you must have the Short Message Service SMS or Telocator Alpha numeric Protocol TAP server number and parameter settings of your cell phones and or pagers 1 Make sure you are connected to the site and select the Alarms tab Figure 2 17 Figure 2 17 Setting up text messaging and alarm conditions 2 Under Alarm Notification select the Al...

Страница 35: ...the field labeled Message type the outgoing text message maximum of 32 characters 6 Under Alarm condition select the retry number and interval and define the alarm condition using the Equa tion Builder Defined alarm conditions will appear on the pull down list 7 Click the Apply button or press F9 when you are done to update the module s settings 2 7 Power Conservation 2105Ci If you are using a 210...

Страница 36: ...2105 Interface Module Section 2 Installation and Operation 2 18 Figure 2 18 Wireless power control window ...

Страница 37: ...n Figure 2 19 select the Module Name the equation will affect Choose an equation from the Equation List or click Add and type in a descriptive name Select a condition from the list on the bottom half of the screen and click on Edit Condition or if no conditions are defined click Edit Condition Figure 2 19 Equation Builder for power control Selecting Edit Condition opens up the Condition Builder sc...

Страница 38: ...uired to use this feature Contact the factory for additional information To use the data push capability you need to make sure you are connected to the interface module and then select the Data tab Figure 2 21 Figure 2 21 Data tab Click the Pushed Data button to set up a schedule for the data to be pushed to a Flowlink Pro server Detailed Flowlink instructions are beyond the scope of this manual F...

Страница 39: ...is available for connection to YSI sondes All other sondes can be connected using the universal interface cable See Figure 2 35 for cable information Scan for and activate or remove connected sondes using the Sonde tab Figure 2 22 To choose the parameters measured by each sonde highlight the sonde in the Active Sondes list and click Configure To set up data storage for each parameter use the Jump ...

Страница 40: ...Teledyne Isco model 674 rain gauge and one for non Isco rain gauges See Figure 2 35 for cable information Note If you have ordered any version of the Teledyne Isco Model 677 logging rain gauge Isco rain gauge and 2105 module combination your system includes the rain gauge interface cable part 69 2004 581 Measurement units are selectable on the Rainfall tab s pull down list Figure 2 23 Total rainfa...

Страница 41: ... the Sampler tab Figure 2 24 you can choose to keep the sampler enabled or disabled or enable the sampler based on con ditions To keep the sampler enabled once it has been triggered select the Latch check box For more information about sampler enabling and rain events refer to your Flowlink Help menu The 2105 can pace the sampler based on total flow volume Be sure that the sampler s programmed sam...

Страница 42: ... can start a program modify the current program or stop an active program This feature is helpful for applications such as storm monitoring where the need to adjust program operation due to pending events is often necessary Note If you use this feature it is important that you follow the steps in Table 2 2 Make sure you close all active menus and close the terminal emulation connection to the samp...

Страница 43: ...connect port of the sampler Figure 2 26 Cable connections for sampler Pass Thru Mode A terminal emulation program such as HyperTerminal is also required In our examples that follow the configuration screen shots are from HyperTerminal but the process and command lines would be similar for any terminal emulation program Open a HyperTerminal session enter a Name and click OK Figure 2 27 The next scr...

Страница 44: ...lick Modify and enter phone number of the site you want to access and select the Settings tab Figure 2 28 Figure 2 28 Specifying the connection number From the Settings tab Figure 2 29 select ANSI for emulation mode and click OK Figure 2 29 HyperTerminal Settings tab ...

Страница 45: ...ou see a response This logs into the 2105 2 SP CR Accesses the System Process menu 3 60 CR 60 enters the ADFM settings 4 ADFM Type ACOM Opens the Pass thru connection 5 Hold the shift key and press the key repeatedly until you see a response This logs into the 6712 or Avalanche 6 MENU CR Enters the Sampler Control menu 7 CONTROL CR Accesses the sampler s keypad 8 CR Arrow keys are used to navigate...

Страница 46: ...ument Note Pulse Doppler input and Modbus input see Section 2 14 can not be used simultaneously in the same system The 2105Ci interface module can push Pulse Doppler ensemble data to a Flowlink Pro database server Pulse Doppler support and pushed data functions are activated by selecting the check box on the ADFM tab Figure 2 31 The pushed data function is configured from the Data tab Figure 2 21 ...

Страница 47: ...with one input The RS 485 will require a converter but it can support up to two inputs Teledyne Isco offers an RS 232 to RS 485 converter Note You must set up and configure your monitoring control system and Modbus device s before adding the 2105 module Note Modbus and Pulse Doppler input see Section 2 13 cannot be used simultaneously in the same system From the Modbus tab configure the protocol T...

Страница 48: ...ss between 1 and 247 Enter a connection retry interval in milliseconds Request Timeout and the number of connection attempts before a connection failure is determined Figure 2 33 Modbus configuration window Click Continue to open the Device Parameters window From this window specify the parameter s measured by the device Figure 2 34 Device parameters measured data Address between 1 and 247 List of...

Страница 49: ...urements tab and may be selected and edited for data storage like any other data type Note All 2100 modules have Modbus output capability For an over view of Modbus principles and an explanation of Modbus out put operation as it applies to 2100 modules turn to Section 3 Modbus Protocol and 2100 Output 2 14 1 4200 and 6700 Series Measured Data This section is specific to 4200 Series flow meters and...

Страница 50: ... screen scroll and select Serial Output Select YES 3 Under Port Settings choose 2400 or 9600 must match the baud rate selected for the 2105 4 Select the desired output interval between 15 seconds and 15 minutes Configuring the 2105 for 4200 6700 To configure the 2105 for interfacing with the 4200 6700 1 Connect the 2105 to the instrument s interrogator port Contact the factory to order the appropr...

Страница 51: ...ent day s rain tips since midnight Tips CR 21 Previous day s rain tips since midnight Tips PR 23 pH pH units PH 25 Dissolved Oxygen Milligrams per liter DO 27 Temperature Degrees Celsius TE 29 YSI 600 pH pH units YPH 31 YSI Dissolved Oxygen Milligrams per liter YDO 33 YSI 600 Conductivity Millisiemens per centimeter YCO 35 YSI 600 Specific Conductance Millisiemens per centimeter YSP 37 YSI 600 Sal...

Страница 52: ...7 Level Meters LE0 89 Ammonium nitrogen Milligrams nitrogen per liter MM0 91 Ammonia nitrogen Milligrams nitrogen per liter MA0 93 Nitrate Nitrogen Milligrams nitrogen per liter NI0 95 Turbidity Nephelometric turbidity units TB0 97 Chloride Milligrams per liter CL0 99 Chlorophyll Micrograms per liter CP0 101 Flow rate Cubic meters per second FL0 103 Flow volume Cubic meters VO0 105 Conductivity Mi...

Страница 53: ... Degrees Celsius TE2 153 Oxidation reduction Potential millivolts OR2 155 Level Meters LE2 157 Ammonium nitrogen Milligrams nitrogen per liter MM2 159 Ammonia nitrogen Milligrams nitrogen per liter MA2 161 Nitrate Nitrogen Milligrams nitrogen per liter NI2 163 Turbidity Nephelometric turbidity units TB2 165 Chloride Milligrams per liter CL2 167 Chlorophyll Micrograms per liter CP2 169 Flow rate Cu...

Страница 54: ...er MMN N 34 91 Ammonia nitrogen Milligrams nitrogen per liter MAN N 17 93 Nitrate Nitrogen Milligrams nitrogen per liter NIN N 17 95 Turbidity Nephelometric turbidity units TBN N 17 97 Chloride Milligrams per liter CLN N 17 99 Chlorophyll Micrograms per liter CPN N 17 101 Flow rate Cubic meters per second FLN N 17 103 Flow volume Cubic meters VON Table 2 3 4200 6700 Series Supported Data Types Con...

Страница 55: ...d associated equipment 600 YSI 674 69 2004 580 Universal Interconnect Cable 10 2105 69 2004 581 Rain Gauge Interconnect Cable 50 2105 69 2004 582 SDI 12 YSI Interconnect Cable 6 2105 69 2004 583 Isco 674 Rain Gauge Interconnect Cable 10 2105 69 2004 584 Y Cable 2105 to Sampler 25 69 2004 589 Y Cable 2105 to 9 pin metal ADFM Canister 25 69 2004 585 Y Cable 2105 Connector 10 69 2004 586 Y Cable 2105...

Страница 56: ...ically insulate any unused wires 3 Ensure that the connection has protection from the envi ronment such as cable conduit Teledyne Isco offers a sealed in line cable joiner for water protection WARNING Ensure that no wire ends come into contact with each other Personal injury power shutdown or component failure can result Wires not in use should be individually sealed at all times 2 16 2 2105 Conne...

Страница 57: ...nsmits sustained 0 VDC for Inhibit F A2D IO5F Blue I O Analog Voltage Sampler Event Mark Input Receives 3 second 4 to 26 VDC pulses G CTR IO6G Violet I O Analog Voltage Rain Gauge Tip Input Receives 48 ms 0 VDC Rain pulses H CTR IO7H Gray I O Counter Sampler Bottle Number Input Receives 48 ms 4 to 26 VDC pulses J DGND J Black Digital Ground Digital Ground 0 VDC 100 mA typical K VBAT K Tan Battery ...

Страница 58: ...2105 Interface Module Section 2 Installation and Operation 2 40 ...

Страница 59: ...a from a single module or stack of modules at a site or multiple sites over a wide area The data can be sent to a central com puter for display data collection or process control Modbus output implementation is independent of Flowlink and cannot alter the Flowlink programmed configuration of the module Modbus cannot be used to retrieve historical data from a module s memory Due to the wide variety...

Страница 60: ...ses a polling mechanism to repeatedly send a command until a response is received It may take up to 20 command retries before the module has identified the baud rate and a response is received Modbus Protocol Setup The communication settings for Modbus protocol are as follows Baud 9600 Bits 8 Parity None Stop Bits 1 3 2 2 Module Addressing When connecting to a site via a Modbus OPC server you use ...

Страница 61: ...vidual 2100 sites through Modbus using the COM ports on the OPC Server which are directly connected to the remote sites Connection to the module is made through the RS 232 communi cation port on the top of the module Note For low power operation we recommend connecting the mod ule s to the computer using the straight through cable Isco part number 60 5314 529 which consumes less power instead of o...

Страница 62: ...h another For output 2100 modules support Modbus ASCII pro tocol Dedicated Line A telecommunications path reserved for com munication between two specified points and not shared among multiple points Modbus Protocol Modbus Protocol is a messaging structure used to establish master slave client server communications between intelligent devices Modbus is a simple command response mechanism to read f...

Страница 63: ... Control PLC Programmable Logic Controller RTU Remote Terminal Unit SCADA Supervisory Control And Data Acquisition TCP IP Transmission Control Protocol Internet Protocol 3 6 RegisterSpecifications 2100 Output All numbers in the Modbus registers are stored most significant byte first If the polling device has a byte ordering of least signif icant byte first an Intel based PC for example the bytes w...

Страница 64: ...N 1 Register Definitions Register Number s Name Data Type Units Read Write 1 4 Model number 8 byte string None Read 5 23 Module name 38 byte string None Read 24a Identify module 16 bit integer None Read Write 25b Take reading flag 16 bit integer None Read Write 26c Update interval 16 bit integer Seconds Read Write 27d Active flag 1 16 bit field None Read 28 Active flag 2 16 bit field None Read 29 ...

Страница 65: ...ime record Time Read 190 191 Analog channel 2 4 byte float 0 100 percent Read 192 Analog channel 2 status code 16 bit integer Read 193 198 Analog channel 2 time Record Time Read 205 206 Analog channel 3 4 byte float 0 100 percent Read 207 Analog channel 3 status code 16 bit integer Read 208 213 Analog channel 3 time record Time Read 220 221 Analog channel 4 4 byte float 0 100 percent Read 222 Anal...

Страница 66: ... Dissolved Gas 4 byte float mmHg Read 597 Dissolved Gas status code 16 bit integer Read 598 603 Dissolved Gas time record Time Read 655 656 Circulator 4 byte float TBD Read 657 Circulator status code 16 bit integer Read 658 663 Circulator time record Time Read 715 716 Photosynthetic Radiation 4 byte float umol s1 m2 Read 717 Photosynthetic Radiation status code 16 bit integer Read 718 723 Photosyn...

Страница 67: ...g l Read 1257 Ammonium Nitrogen status code 16 bit integer Read 1258 1263 Ammonium Nitrogen time record Time Read 1315 1316 Nitrate Nitrogen 4 byte float mg l Read 1317 Nitrate Nitrogen status code 16 bit integer Read 1317 1323 Nitrate Nitrogen time record Time Read 1375 1376 Turbidity 4 byte float NTU Read 1377 Turbidity status code 16 bit integer Read 1378 1383 Turbidity time record Time Read 14...

Страница 68: ...5 second update of readings when a Modbus mas ter is polling the 2100 c The Update Interval specifies an interval in seconds that the registers are automatically updated It defaults to zero which indicates that no automatic updating will occur d The Active Flag 1 4 bit fields specify what fields registers are active in the list This provides support for a maxi mum of 64 fields For example if bit 0...

Страница 69: ...e 3 3 Measurement Parameters by Model Number 2103 2103c g 2105 2105c g 2108 2110 2150 2151 2151P Voltage Analog channel 1 Level Level Analog channel 2 Flow Velocity Analog channel 3 Volume Flow Voltage Flow 1 Temperature Volume Volume 1 Voltage Temperature Subject to change ...

Страница 70: ...2105 Interface Module Section 3 Modbus Protocol and 2100 Output 3 12 ...

Страница 71: ... of the module s operation under dif fering environmental conditions a weekly maintenance interval is recommended 4 1 1 Cleaning The module case may be cleaned using a soft cloth warm water and a mild detergent Do not use an abrasive cleanser or you might scratch the surface of the case Before cleaning make sure that all the protective connector caps are in place to avoid damage to any of the conn...

Страница 72: ...nted oven in a well ventilated room Do not remain in the room while the regeneration is taking place Use the recommended temperature Avoid heating the desiccant at higher than recommended temperatures There is the potential of irritating fumes coming from the des iccant during reactivation Because of this we urge you to use caution and to heat the desiccant in a well ventilated room Material Safet...

Страница 73: ...ound the circumference of the ring Note Do not use petroleum based lubricants Petroleum based lubricants will cause the O ring to swell and eventually deterio rate Aerosol silicone lubricant sprays often use petroleum based propellants If you are using an aerosol spray allow a few minutes for the propellant to evaporate before proceeding 4 5 How to Obtain Service The internal components of the 210...

Страница 74: ...2105 Interface Module Section 4 Maintenance 4 4 ...

Страница 75: ...in this section Reference the call outs in the accompanying tables to determine the part number for the item Replacement parts can be purchased by contacting Teledyne Isco s Customer Service Department Teledyne Isco Customer Service Department P O Box 82531 Lincoln NE 68501 USA Phone 800 228 4373 402 464 0231 FAX 402 465 3022 E mail IscoInfo teledyne com ...

Страница 76: ...2105 Interface Module Appendix A Replacement Parts List A 2 A 1 1 2105 Interface Module Drawing 602002586 ...

Страница 77: ...2105 Interface Module Appendix A Replacement Parts List A 3 Drawing 602002586 ...

Страница 78: ...2105 Interface Module Appendix A Replacement Parts List A 4 Drawing 602002586 ...

Страница 79: ...2105 Interface Module Appendix A Replacement Parts List A 5 ...

Страница 80: ...2105 Interface Module Appendix A Replacement Parts List A 6 ...

Страница 81: ...2105 Interface Module Appendix A Replacement Parts List A 7 A 1 2 2191 Battery Module Drawing 602003113 ...

Страница 82: ...2105 Interface Module Appendix A Replacement Parts List A 8 ...

Страница 83: ...2105 Interface Module Appendix A Replacement Parts List A 9 A 1 3 2105Ci Antenna Drawing 602002583 ...

Страница 84: ...2105 Interface Module Appendix A Replacement Parts List A 10 ...

Страница 85: ...dyne com B 2 General Accessories 2105 2015Ci Instruction Manual 69 2003 635 Alkaline Lantern Battery 340 2006 02 Rechargeable 6V Lead acid Lantern Battery 60 2004 041 Charging Adapter for 6V Lead acid Lantern Battery 60 2004 040 Flowlink and Flowlink Pro Software call factory Isco Open Channel Flow Measurement Handbook 60 3003 041 Magnetic Mount Dual Band Antenna 60 2004 564 CAUTION Only use this ...

Страница 86: ... 2004 507 Universal Interconnect Cable 69 2004 580 non terminated Rain Gauge Cable 69 2004 581 dual spade connectors SDI 12 YSI Cable 69 2004 582 Rain Gauge Model 674 Cable 69 2004 583 4200 Series Flow Meter Cable 60 5314 855 Y Cable 2105 Isco Sampler 69 2004 584 Y Cable 2105 Connectors 69 2004 585 Y Cable 2105 4700 Sampler 69 2004 586 Y Cable 2105 ADFM Box AccQmin 69 2004 587 Y Cable 2105 ADFM Ca...

Страница 87: ...dix to the manual provides Material Safety Data Sheets for the desiccant used by the 2105 Module Teledyne Isco cannot guarantee the accuracy of the data Specific questions regarding the use and handling of the products should be directed to the manufacturer listed on the MSDS ...

Страница 88: ...20 1 2 1 Vapor Pressure mm Hg and Temperature N A Melting Point N A Vapor Density Air 1 N A Evaporation Rate __________ 1 N A Solubility in Water Insoluble but will adsorb moisture Water Reactive Not reactive but will adsorb moisture Appearance and Odor Purple crystals no odor Section 3 Fire and Explosion Hazard Data Flash Point and Methods Used N A Auto Ignition Temperature N A Flammability Limit...

Страница 89: ... at least 2 glasses of water Section 6 Control and Protective Measures Respiratory Protection Specify Type Use NIOSH approved dust mask or respirator Protective Gloves Light cotton gloves Eye Protection Safety glasses VENTILATION Local Exhaust Mechanical General Special TO BE USED Other Specify Other Protective Clothing and Equipment None Hygienic Work Practices Avoid raising dust Avoid contact wi...

Страница 90: ...re not combustible Potential Health Effects Eyes None Skin None Ingestion Ingestion is very unlikely but if ingested get medical attention Inhalation None Medical Effects Generally Aggravated by Exposure None Chronic Effects Carcinogenity None Section 3 Composition Information on Ingredients Component Name CAS Number by Weight Synthetic amorphous silica gel SiO2 112926 00 8 80 90 Cobalt chloride 7...

Страница 91: ...te disposal container Avoid raising dust Wash with soap and water after handling Section 7 Handling and Storage Handling Avoid raising dust and minimize the contact between worker and the material Practice good hygienic work practices Storage Store in a cool dry location Keep in sealed containers away from moisture The silica gel will readily adsorb moisture Section 8 Exposure Controls Personal Pr...

Страница 92: ...icity in Humans ACGIH Cobalt and inorganic compounds as Co are rated as A3 confirmed animal carcinogen and unknown relevance to humans IARC Cobalt and cobalt compounds are rated as Group 2B possible carcinogens to humans Human Toxicology Silica gel is a synthetic amorphous silica not to be confused with crystalline silica Epidemiological studies indicate low potential for adverse health effects In...

Страница 93: ...t place HMIS PPE designation codes on the MSDSs or labels that leave the facility as it is not known the conditions under which the customer will use this product This MSDS was prepared by George E Mckedy Senior Applications Development Specialist Multisorb Technologies Inc This data and recommendations presented in this data sheet concerning the use of our product and the materials contained ther...

Страница 94: ...2105 Interface Module Appendix C Material Safety Data Sheets C 8 ...

Страница 95: ...y 2 20 rain gauge interface 2 22 sampler interface 2 23 SDI 12 input 2 21 setting up text messaging 2 16 Flowlink software 2 9 I installation 2 3 batteries 2 6 connectors 2 5 latches 2 4 L latches 2 4 stacking modules 2 7 M maintenance 4 1 material safety data C 1 Modbus 4200 6700 data input 2 31 input 2 29 output 3 1 registers 3 5 modem connection 2 12 CSD 2 15 serial over IP 2 14 modem module po...

Страница 96: ...2105 Interface Module Index Index 2 service 4 3 sondes 2 21 modbus registers 3 8 specifications 1 6 stacking 2 7 T TCP connection 2 14 technical specifications 1 6 text messaging 2 16 ...

Страница 97: ...States if the instrument proves to be defective within 30 days of shipment Throughout the remainder of the warranty period the customer will pay to return the instrument to Teledyne Isco and Teledyne Isco will pay surface transportation to return the repaired instrument to the customer Teledyne Isco will not pay air freight or customer s packing and crating charges This warranty does not cover los...

Страница 98: ......

Отзывы: